Service Manual MA 42
D-0125965 Rev. D
16
18/05/2022
4.2 Software Calibration
For the software calibration of the MA42 the following is required.
1. MA42 to a PC with a USB 2.0 cable.
2. MA42 Calibration Tool.
3. Service Code available for approved service partners
4.2.1 Installation of Calibration Tool
Approved service partners will receive the MA42MA52 Calibration.msi file
Step 1: Run msi file with a double click: Windows Installer will open up and prepare the
installation.
From here you follow all the prompts that are requested. Press to continue or abort
with Cancel.