Appendix A Specification
Page A-2 MAN 0385
A
Hydro 2000S
Purge air specification (Hydro 2000G)
This section only applies to the Hydro 2000G.
If a particularly aggressive dispersant is used, this may cause the grease in the pump
motor bearings to leach out. If this may be a problem the air purge inlet port can be
used to connect an air supply. This will keep fumes away from the bearings.
If required to use the air purge inlet connector (see Chapter 2), check that the air
supply conforms to the following specification:
1800ml air/min at 212mbar g.
0.064 cu ft/min at 3.0psi g.
Filtered to remove all particles larger than 0.5 microns (if the dispersion unit is
to be used in unclean environments).
Dispersion type Wet
Capacity 50-150ml
Typical applications Solvent-based suspensions, Pharmaceuticals.
Dispersion mechanisms Continuously variable combined pump/stirrer.
Continuously variable ultrasonics
Modes of operation Automatic via SOPs.
Manual via computer operating dialogues.
Weight 16.8kg
Dimensions Width: 348mm
Height: 333mm
Depth: 365mm
Power 100-120V / 200/240 V AC, 50/60Hz, 240VA
Maximum size of particles Up to 600 microns, depending on particle shape
and density.
MAN0385-1.0 Hydro 2000G-S.book Page 2 Friday, March 16, 2007 4:02 PM