What causes the Maquet Medical Equipment lighthead to drift downwards?
- HHeather DiazSep 2, 2025
If the lighthead of your Maquet Medical Equipment drifts downwards, the spring-balance in the spring-arm is likely worn. Readjust the spring-balance.
What causes the Maquet Medical Equipment lighthead to drift downwards?
If the lighthead of your Maquet Medical Equipment drifts downwards, the spring-balance in the spring-arm is likely worn. Readjust the spring-balance.
Why is the Maquet Medical Equipment light not white or the color is wrong?
If the Maquet Medical Equipment light isn't white or the color is incorrect, it could be due to the wrong bulb(s) or lens. Replace the bulb and check the lens version.
How to fix a Maquet HANAULUX 2000 light that is heavy and difficult to move?
If your Maquet Medical Equipment light is heavy and difficult to move, the brakes may be too tight. Try readjusting the brakes.
What to do if the Maquet Medical Equipment light is too light and drifts easily?
If your Maquet Medical Equipment light drifts easily, the brakes are likely too loose or worn. Try readjusting the brakes or replacing them.
What causes Maquet HANAULUX 2000 Medical Equipment bulbs to explode?
Exploding bulbs in Maquet Medical Equipment can be attributed to excessively high voltage or a short circuit in the wiring. An electrician should check the equipment.
Why is the bulb life so short in my Maquet HANAULUX 2000 light?
A short bulb life in your Maquet Medical Equipment light might be caused by using the wrong bulb or having voltage that is too high. Check the bulb and replace it if needed. An electrician should also check the voltage.
What to do if the Maquet Medical Equipment light is too bright?
If the Maquet Medical Equipment light is too bright, it could be due to incorrect bulbs or excessively high voltage. You should check the bulb and replace it if necessary. Also, an electrician should check the voltage.
Why is the Maquet Medical Equipment light intensity too low?
Several factors could cause low light intensity in your Maquet Medical Equipment light. These include low voltage, incorrect bulbs, damaged lenses from improper cleaning agents, or using the wrong lens. To resolve this: an electrician should check the voltage, verify and replace the bulb if necessary, replace the lenses and change the cleaning agent, and check the lens version.
Brand | Maquet |
---|---|
Model | HANAULUX 2000 |
Category | Medical Equipment |
Language | English |
Essential precautions for safe operation and potential hazards.
Defines the intended applications and limitations of the HANAULUX 2000 system.
Lists critical conditions for ensuring the luminaire system's safe and correct function.
Specifies the recommended temperature, humidity, and pressure for transport and storage.
Provides guidance on how to dispose of old units and components responsibly.
Explains the modular design, flexibility, and core principles of the HANAULUX 2000 lighting system.
Lists and identifies the standard, interchangeable components that make up the HANAULUX 2000 system.
Introduces specialized parts designed to enhance system functionality and adaptability for specific needs.
Explains versions with restricted rotation due to integrated cables and reasons for such limitations.
Outlines the system's capability to be prepared for future additions like SATELLIT lights or cameras.
Details the HANAUCHROME optical system, bulb technology, and reflector design that ensure cool, high-quality illumination.
Explains how the Fresnel lens creates a deep, homogeneous light field with excellent operating depth.
Highlights the sealed housing, smooth surfaces, and chemical-resistant coatings for easy cleaning and hygiene.
Describes features like graduated arms, spacers, and joint brakes that ensure smooth, precise light positioning.
Provides step-by-step instructions for attaching and detaching the removable, sterilisable light handle.
Explains how to balance and adjust the spring arm for the HANAUPORT Lite accessory with mounted equipment.
Guides on how to position the lightheads using the handles, including specific arm configurations (M position).
Introduces the process of adjusting the light field size and focus for optimal illumination.
Advises on setting the minimum field size to prevent glare and eye strain in the operating area.
Defines the optimal working distance and the depth of the focused light column for consistent illumination.
Recommends using a dimmer and provides guidance on adjusting light intensity for operations.
Covers specific operational features for the 2007 iXL D and 2007 iXL AF models.
Explains the autofocus system, its range, activation, and deactivation for the 2007 IXL AF.
Details how to adjust light intensity using the touch panel controls on specific models.
Provides essential information and warnings regarding the specialized halogen bulbs for the HANAULUX 2000 system.
Explains how to adjust brake screws for proper lighthead movement and stable positioning, noting model variations.
Details how to readjust spring balances when lightheads drift, varying by spring-arm version.
Step-by-step instructions for adjusting spring arms on specific HANAULUX 2002 i to 2005 i/iXL models.
Instructions for adjusting the spring force on straight spring-arm versions for optimal balance.
Specific procedures for adjusting the upper stop and balancing of the Acrobat 2000 standard spring arm.
Specific procedures for adjusting the upper stop and balancing of the Acrobat 2000 straight spring arm.
Guidelines for cleaning and disinfecting the light unit's special coating and chrome-plated handles.
Recommendations for cleaning the tough acrylic lenses, advising against alcohol-based agents.
Specifies the correct procedure and conditions for sterilising the handles to ensure durability.
Refers to a separate manual for specific cleaning instructions related to the HANAUVISION camera system.
Lists routine checks operators should perform every six months to identify potential issues.
Outlines the annual servicing requirements and checks performed by authorized service personnel.
Provides guidance on identifying and ordering correct replacement parts like reflectors and lenses.
Addresses issues related to heavy movement, drifting lights, and spring-arm problems.
Covers problems with light flickering, intensity, color accuracy, and field roundness.
Details troubleshooting steps for cracked, split, melted, or bubbled lenses due to cleaning or bulb issues.
Provides solutions for handles not lasting long, being discolored, or falling off unexpectedly.
Addresses issues where the light does not focus correctly or the focus mechanism is stiff.
Presents detailed technical specifications for light intensity, field size, working area, and focus range across models.
Lists mechanical specifications such as rotation diameter, vertical action, and weight for various HANAULUX models.