1.5 Environment operating conditions
The instrument should be protected from conditions that could cause exposure to frost, dew, percolating
water, rain, excessive direct sunlight, etc.
Performance can be affected by sources of heat and cold from heating, air conditioning systems, or drafts.
It is advisable to operate the system in a clean laboratory environment, with minimal atmospheric
concentrations of organic vapours.
1.5.1 Temperature
Recommended operating ambient temperature range is 15 to 30°C.
1.5.2 Humidity
Recommended operating humidity range is 5 to 95% non-condensing.
1.5.3 Altitude
This product should not be operated above 2000m (~6500ft ).
NOTE: For storage or shipping the allowable temperature range is -40 to 70°C and the allowable humidity
range is 5-95% non-condensing. After instrument exposure to extremes of temperature or
humidity, allow 2 hours for return to the recommended ranges before switching on.
1.6 Technical Specifications
1.6.1 Physical properties
1.7 Technical support contact details
In the first instance please contact your distributor. If they are unable to resolve your query, please contact
Markes International (details below).
Website: www.markes.com
E-Mail: support@markes.com
Telephone: +44 (0)1443 230935 (UK office)
+49 (0)6102 8825569 (German Office)
+1 866 483 5684 (US office (toll-free))