MAINTENANCE
Cleaning
DAILY CLEANING PROCEDURE (AT THE END OF
EACH DAY):
When sterilizing culture mediums that generate sul-
de gas or chlorine gas, the inside of the chamber
must be cleaned and rinsed thoroughly without fail.
1. Remove bottom splash bafe.
IMPORTANT
STERILIZING CHAMBER MUST BE CLEANED
AND DRAINED DAILY USING THE FOLLOW-
ING PROCEDURE. WASH WETTED PORTION
OF THE CYLINDER THOROUGHLY BY ADDING
A MILD DETERGENT TO WATER IN CYLIN-
DER.
2. If a soft cloth or brush is used with the detergent and
does not completely remove the surface lm, a nylon
soap pad should be used. After washing thoroughly
rinse with clean water. Dry cylinder* and leave door
open overnight.
* The Sterilmatic cylinder is constructed of corrosion re-
sistant Alclad aluminum alloy. The protective properties of
this material afforded to the interior portion of the cylinder
which is exposed to water may be destroyed by allowing
a lm to form. Such a lm can be caused by salts or other
contaminants in the water. Corrosion may also occur if
water is not drained daily.