Serial Input Description
A mA measurement
a mA source
I mA loop
i mA 2W Sim
V Volts measurement
v Volts source
M mV measurement
m mV source
E HVAC measure
e HVDC measure
K Khz measurement
k Khz source
H Hz measurement
h Hz source
P CPM measurement
p CPM source
W 2-wire measurement (Ohms and RTDs)
X 3-wire measurement (Ohms and RTDs)
Y 4-wire measurement (Ohms and RTDs)
T Thermocouple measurement (default Type J) use
"S" command to select sensor type
t Thermocouple source (default Type J) use "S"
command to select sensor type
C Selects Centigrade ( T/C - RTD )
F Selects Fahrenheit ( T/C - RTD )
R RTD measurement mode (default Pt100 385) use
"S" command to select sensor type
r RTD measurement mode (default Pt100 385) use
"S" command to select sensor type
0-9 Enter a source value using ascii characters
-,. 0,1,2,..,9,-,.terminated by <CR>(carriage
<CR> return)
D Single Broadcast of most recent display value and
units
d Single Broadcast of most recent display value
"S" Commands Select
Sensor Type
Serial Input Selection Entry Description
S No. T/C Type RTD Type
1 J Pt100 (3926)
2 K Pt100 (385)
3 T Pt100 (3916)
4 E Cu10
5 R Pt200 (385)