12 www.mastercool.com
THERMOMÈTRE INFRAROUGE MULTIFONCTION MÉDICAL/SURFACE
SPÉCIFICATIONS:
• Plage de mesure de la température:
Mode front: 34 - 42,2°C (93,2 - 108°F)
Mode surface: -22 - 80°C (-7,6 - 176°F)
• Plage de température de fonctionnement:
10 - 40°C (50 - 104°F), 15 - 85% d’humidité relative
• Plage de température de stockage: stocké à température ambiante
entre -20 à - 50°C (-4 et 122°F), H.R. <85%
• La température de transport doit être inférieure à 70°C (158°F),
H.R. <95%
• Pression atmosphérique: 800 - 1013 hPa
• Précision:
Mode front : ±0,2°C (0,4°F) entre 95 et 107,6 °F (35 - 42°C)
(Température ambiante: 15 - 35°C (59 - 95°F), ±0,3°C (0,5°F)
pour autre plage mode Surface: ±0,3°C (0,5°F) entre 22 et
42,2°C 71,6 et 108°F), autres ±4 % ou ±2°C (4°F), la valeur la plus élevée étant retenue
• Pile: 2 de type AAA
• Durée de vie des piles: environ 3 000 lectures en continu
• Durée de vie prévue: 4 ans
• Rétro-éclairage à LEDs bleues: s’allume automatiquement après la mesure et s’éteint automatiquement après 5
secondes
• Étanchéité: IP22
REMARQUE : Le thermomètre est étalonné durant la fabrication. Si vous avez des questions sur le mode d’étalonnage,
la précision des mesures de température ou des événements imprévus à tout moment, veuillez contacter le service
d’assistance.
Avertissement : Aucune modification de cet équipement n’est autorisée.
Avertissement : Ne pas immerger l’appareil dans des liquides et ne pas l’exposer à l’humidité directe.
FONCTIONS
TEMPÉRATURE DU FRONT:
Ce thermomètre n’a pas pour but de remplacer une visite chez le médecin. Ne pas oublier non plus
de comparer le résultat de la mesure par rapport à votre température corporelle habituelle. Veuillez
consulter un médecin si vous avez des problèmes de santé.
• Veuillez consulter la section « UTILISATION DU THERMOMÈTRE » pour savoir comment mesurer la
température du corps.
THD2FE Non-contact Forehead Thermometer
Functions
The thermometer has been designed for home use. It’s not meant to replace a visit to the
doctor. Please also remember to compare the measurement result to your regular body
temperature. Please consult with doctor if you have health concerns.
à Please see the “Use of the thermometer”
section to learn how to measure the body
temperature.
The surface mode shows the actual and unadjusted surface temperature which is different from
the body temperature. It can help you monitor if the object temperature is suitable for the baby
or patient, for example the baby’s milk.
à Please see the “Use of the thermometer” section to learn how to measure
the object
temperature.
If the thermometer detects a temperature ≧99.5°F (or 37.5°C) under forehead mode, three
short beep sound will follow one long beep sound to warn the user for the temperature is above
99.5°F (37.5°C).
There are total 25 sets of measurement condition for body temperature.
à When power on, press the “ON/MEM” button to see the temperature records with icon.
In “Power Off” mode, press and hold the “START” button, then press the “ON/MEM” button for
3 seconds, icon “°F” will be switched to icon “°C”. You can also use the same process to
change the LCD display from °C to °F.
NOTE:
Memory clear and Celsius/Fahrenheit switch are together. When you switch Celsius and
Fahrenheit, the Memory will be cleared.
The device setting with buzzer is on, you can set buzzer on/off under Mute mode.
When power on, press and hold the “ON/MEM” button for 3 seconds. The icon
will flash on
the LCD screen and then release the “ON/MEM” button to set MUTE. Thus you will not hear
beep sounds. You can also use the same process to turn off the Mute function.
NOTE: If keep pressing "ON/MEN" button for 2 seconds after
icon flashing, the device will
be power off WITHOUT setting Mute.
Use of the thermometer
Note: If there is any temperature difference between the places where the device is stored and where you are going to measure, subject and the device should
stay in the same room for at least 15 minutes before measurement.
1. Always make sure the probe lens is clean without any damage and the forehead is clean.
Warning: Choking from swallowing small parts and batteries by children or pets is possible, please keep small parts and batteries at places where
children and pets can’t reach.
2. Power on:
Press the “ON/MEM” button (see figure 1).
3. Measuring body temperature on the forehead:
The Non-contact Clinical Thermometer, Model THD2FE is an infrared thermometer intended
for the intermittent measurement of human body temperature in people of all ages.
At least 11 years old (5 years intensive reading experience), no maximum.
þ Temperature measurement range:
Forehead mode: 93.2~108°F (34~42.2°C)
Surface mode: -7.6~176°F (-22~80°C)
þ Operating temperature range: 50~104°F (10~40°C), 15%~85% RH
þ Storage temperature range: It should be stored at room temperature between -4~122°F
(-20~+50°C), RH≦85%
Transportation temperature shall be less than 158°F (70°C), RH≦95%
þ Atmospheric pressure: 800~1013 hPa
þ Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC),
IEC/EN60601-1(Safety) standards, ISO10993, RoHS.
þ Accuracy:
Forehead mode: ±0.4
o
F (0.2
°C ) within 95~107.6°F (35~42°C),
±0.5
o
F (0.3
°C ) for other range.
Surface mode: ±0.5°F (0.3°C) within 71.6~108°F (22~42.2°C),
others ±4% or ±4°F (2°C) whichever is greater.
þ Temperature indicator, memory and °F / °C switch function
þ Battery: AAA x 2 pcs
þ Battery life: around 3,000 continuous readings.
þ Expected Service Life: 4 years
þ This thermometer converts the forehead temperature to display its “oral equivalent.”
(according to the result of the clinical evaluation to get the offset value)
þ Blue LED Backlight
Blue LED backlight will be automatically turned on after measurement, and automatically
turned off after 2 seconds.
þ Enclosure Rating: IP22
þ Dimensions: 158.0 x 48.0 x 40.2 mm
þ Weight: 100 grams including battery
The device should not submerge into any liquids and expose it to direct moisture.
There is no gender and age limitation for using the infrared thermometer.
This is not an AP or APG product.
TEMPÉRATURE DE SURFACE:
Le mode surface indique la température de surface réelle et non ajustée qui est différente de la
température du corps.
• Veuillez consulter la section « UTILISATION DU THERMOMÈTRE » pour savoir comment mesurer la
température d’un objet.
THD2FE Non-contact Forehead Thermometer
Functions
The thermometer has been designed for home use. It’s not meant to replace a visit to the
doctor. Please also remember to compare the measurement result to your regular body
temperature. Please consult with doctor if you have health concerns.
à Please see the “Use of the thermometer”
section to learn how to measure the body
temperature.
The surface mode shows the actual and unadjusted surface temperature which is different from
the body temperature. It can help you monitor if the object temperature is suitable for the baby
or patient, for example the baby’s milk.
à Please see the “Use of the thermometer” section to learn how to measure
the object
temperature.
If the thermometer detects a temperature ≧99.5°F (or 37.5°C) under forehead mode, three
short beep sound will follow one long beep sound to warn the user for the temperature is above
99.5°F (37.5°C).
There are total 25 sets of measurement condition for body temperature.
à When power on, press the “ON/MEM” button to see the temperature records with icon.
In “Power Off” mode, press and hold the “START” button, then press the “ON/MEM” button for
3 seconds, icon “°F” will be switched to icon “°C”. You can also use the same process to
change the LCD display from °C to °F.
NOTE:
Memory clear and Celsius/Fahrenheit switch are together. When you switch Celsius and
Fahrenheit, the Memory will be cleared.
The device setting with buzzer is on, you can set buzzer on/off under Mute mode.
When power on, press and hold the “ON/MEM” button for 3 seconds. The icon
will flash on
the LCD screen and then release the “ON/MEM” button to set MUTE. Thus you will not hear
beep sounds. You can also use the same process to turn off the Mute function.
NOTE: If keep pressing "ON/MEN" button for 2 seconds after
icon flashing, the device will
be power off WITHOUT setting Mute.
Use of the thermometer
Note: If there is any temperature difference between the places where the device is stored and where you are going to measure, subject and the device should
stay in the same room for at least 15 minutes before measurement.
1. Always make sure the probe lens is clean without any damage and the forehead is clean.
Warning: Choking from swallowing small parts and batteries by children or pets is possible, please keep small parts and batteries at places where
children and pets can’t reach.
2. Power on:
Press the “ON/MEM” button (see figure 1).
3. Measuring body temperature on the forehead:
The Non-contact Clinical Thermometer, Model THD2FE is an infrared thermometer intended
for the intermittent measurement of human body temperature in people of all ages.
At least 11 years old (5 years intensive reading experience), no maximum.
þ Temperature measurement range:
Forehead mode: 93.2~108°F (34~42.2°C)
Surface mode: -7.6~176°F (-22~80°C)
þ Operating temperature range: 50~104°F (10~40°C), 15%~85% RH
þ Storage temperature range: It should be stored at room temperature between -4~122°F
(-20~+50°C), RH≦85%
Transportation temperature shall be less than 158°F (70°C), RH≦95%
þ Atmospheric pressure: 800~1013 hPa
þ Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC),
IEC/EN60601-1(Safety) standards, ISO10993, RoHS.
þ Accuracy:
Forehead mode: ±0.4
o
F (0.2
°C ) within 95~107.6°F (35~42°C),
±0.5
o
F (0.3
°C ) for other range.
Surface mode: ±0.5°F (0.3°C) within 71.6~108°F (22~42.2°C),
others ±4% or ±4°F (2°C) whichever is greater.
þ Temperature indicator, memory and °F / °C switch function
þ Battery: AAA x 2 pcs
þ Battery life: around 3,000 continuous readings.
þ Expected Service Life: 4 years
þ This thermometer converts the forehead temperature to display its “oral equivalent.”
(according to the result of the clinical evaluation to get the offset value)
þ Blue LED Backlight
Blue LED backlight will be automatically turned on after measurement, and automatically
turned off after 2 seconds.
þ Enclosure Rating: IP22
þ Dimensions: 158.0 x 48.0 x 40.2 mm
þ Weight: 100 grams including battery
The device should not submerge into any liquids and expose it to direct moisture.
There is no gender and age limitation for using the infrared thermometer.
This is not an AP or APG product.
INDICATEUR DE TEMPÉRATURE:
Si le thermomètre détecte une température >37,5 °C (ou 99,5 °F) en mode front, trois bips courts
retentissent, suivis d’un bip long pour avertir l’utilisateur que la température est supérieure à 37,5 °C
(99,5 °F)
THD2FE Non-contact Forehead Thermometer
Functions
The thermometer has been designed for home use. It’s not meant to replace a visit to the
doctor. Please also remember to compare the measurement result to your regular body
temperature. Please consult with doctor if you have health concerns.
à Please see the “Use of the thermometer”
section to learn how to measure the body
temperature.
The surface mode shows the actual and unadjusted surface temperature which is different from
the body temperature. It can help you monitor if the object temperature is suitable for the baby
or patient, for example the baby’s milk.
à Please see the “Use of the thermometer” section to learn how to measure
the object
temperature.
If the thermometer detects a temperature ≧99.5°F (or 37.5°C) under forehead mode, three
short beep sound will follow one long beep sound to warn the user for the temperature is above
99.5°F (37.5°C).
There are total 25 sets of measurement condition for body temperature.
à When power on, press the “ON/MEM” button to see the temperature records with icon.
In “Power Off” mode, press and hold the “START” button, then press the “ON/MEM” button for
3 seconds, icon “°F” will be switched to icon “°C”. You can also use the same process to
change the LCD display from °C to °F.
NOTE:
Memory clear and Celsius/Fahrenheit switch are together. When you switch Celsius and
Fahrenheit, the Memory will be cleared.
The device setting with buzzer is on, you can set buzzer on/off under Mute mode.
When power on, press and hold the “ON/MEM” button for 3 seconds. The icon
will flash on
the LCD screen and then release the “ON/MEM” button to set MUTE. Thus you will not hear
beep sounds. You can also use the same process to turn off the Mute function.
NOTE: If keep pressing "ON/MEN" button for 2 seconds after
icon flashing, the device will
be power off WITHOUT setting Mute.
Use of the thermometer
Note: If there is any temperature difference between the places where the device is stored and where you are going to measure, subject and the device should
stay in the same room for at least 15 minutes before measurement.
1. Always make sure the probe lens is clean without any damage and the forehead is clean.
Warning: Choking from swallowing small parts and batteries by children or pets is possible, please keep small parts and batteries at places where
children and pets can’t reach.
2. Power on:
Press the “ON/MEM” button (see figure 1).
3. Measuring body temperature on the forehead:
The Non-contact Clinical Thermometer, Model THD2FE is an infrared thermometer intended
for the intermittent measurement of human body temperature in people of all ages.
At least 11 years old (5 years intensive reading experience), no maximum.
þ Temperature measurement range:
Forehead mode: 93.2~108°F (34~42.2°C)
Surface mode: -7.6~176°F (-22~80°C)
þ Operating temperature range: 50~104°F (10~40°C), 15%~85% RH
þ Storage temperature range: It should be stored at room temperature between -4~122°F
(-20~+50°C), RH≦85%
Transportation temperature shall be less than 158°F (70°C), RH≦95%
þ Atmospheric pressure: 800~1013 hPa
þ Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC),
IEC/EN60601-1(Safety) standards, ISO10993, RoHS.
þ Accuracy:
Forehead mode: ±0.4
o
F (0.2
°C ) within 95~107.6°F (35~42°C),
±0.5
o
F (0.3
°C ) for other range.
Surface mode: ±0.5°F (0.3°C) within 71.6~108°F (22~42.2°C),
others ±4% or ±4°F (2°C) whichever is greater.
þ Temperature indicator, memory and °F / °C switch function
þ Battery: AAA x 2 pcs
þ Battery life: around 3,000 continuous readings.
þ Expected Service Life: 4 years
þ This thermometer converts the forehead temperature to display its “oral equivalent.”
(according to the result of the clinical evaluation to get the offset value)
þ Blue LED Backlight
Blue LED backlight will be automatically turned on after measurement, and automatically
turned off after 2 seconds.
þ Enclosure Rating: IP22
þ Dimensions: 158.0 x 48.0 x 40.2 mm
þ Weight: 100 grams including battery
The device should not submerge into any liquids and expose it to direct moisture.
There is no gender and age limitation for using the infrared thermometer.
This is not an AP or APG product.
EMPLACEMENTS DE MÉMOIRE:
TIl existe 25 ensembles de conditions de mesure de la température corporelle.
• Lorsque l’appareil est allumé, appuyer sur le bouton « ON/MEM » pour voir les enregistrements de
température avec l’icône
THD2FE Non-contact Forehead Thermometer
Functions
The thermometer has been designed for home use. It’s not meant to replace a visit to the
doctor. Please also remember to compare the measurement result to your regular body
temperature. Please consult with doctor if you have health concerns.
à Please see the “Use of the thermometer”
section to learn how to measure the body
temperature.
The surface mode shows the actual and unadjusted surface temperature which is different from
the body temperature. It can help you monitor if the object temperature is suitable for the baby
or patient, for example the baby’s milk.
à Please see the “Use of the thermometer” section to learn how to measure
the object
temperature.
If the thermometer detects a temperature ≧99.5°F (or 37.5°C) under forehead mode, three
short beep sound will follow one long beep sound to warn the user for the temperature is above
99.5°F (37.5°C).
There are total 25 sets of measurement condition for body temperature.
à When power on, press the “ON/MEM” button to see the temperature records with icon.
In “Power Off” mode, press and hold the “START” button, then press the “ON/MEM” button for
3 seconds, icon “°F” will be switched to icon “°C”. You can also use the same process to
change the LCD display from °C to °F.
NOTE:
Memory clear and Celsius/Fahrenheit switch are together. When you switch Celsius and
Fahrenheit, the Memory will be cleared.
The device setting with buzzer is on, you can set buzzer on/off under Mute mode.
When power on, press and hold the “ON/MEM” button for 3 seconds. The icon
will flash on
the LCD screen and then release the “ON/MEM” button to set MUTE. Thus you will not hear
beep sounds. You can also use the same process to turn off the Mute function.
NOTE: If keep pressing "ON/MEN" button for 2 seconds after
icon flashing, the device will
be power off WITHOUT setting Mute.
Use of the thermometer
Note: If there is any temperature difference between the places where the device is stored and where you are going to measure, subject and the device should
stay in the same room for at least 15 minutes before measurement.
1. Always make sure the probe lens is clean without any damage and the forehead is clean.
Warning: Choking from swallowing small parts and batteries by children or pets is possible, please keep small parts and batteries at places where
children and pets can’t reach.
2. Power on:
Press the “ON/MEM” button (see figure 1).
3. Measuring body temperature on the forehead:
The Non-contact Clinical Thermometer, Model THD2FE is an infrared thermometer intended
for the intermittent measurement of human body temperature in people of all ages.
At least 11 years old (5 years intensive reading experience), no maximum.
þ Temperature measurement range:
Forehead mode: 93.2~108°F (34~42.2°C)
Surface mode: -7.6~176°F (-22~80°C)
þ Operating temperature range: 50~104°F (10~40°C), 15%~85% RH
þ Storage temperature range: It should be stored at room temperature between -4~122°F
(-20~+50°C), RH≦85%
Transportation temperature shall be less than 158°F (70°C), RH≦95%
þ Atmospheric pressure: 800~1013 hPa
þ Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC),
IEC/EN60601-1(Safety) standards, ISO10993, RoHS.
þ Accuracy:
Forehead mode: ±0.4
o
F (0.2
°C ) within 95~107.6°F (35~42°C),
±0.5
o
F (0.3
°C ) for other range.
Surface mode: ±0.5°F (0.3°C) within 71.6~108°F (22~42.2°C),
others ±4% or ±4°F (2°C) whichever is greater.
þ Temperature indicator, memory and °F / °C switch function
þ Battery: AAA x 2 pcs
þ Battery life: around 3,000 continuous readings.
þ Expected Service Life: 4 years
þ This thermometer converts the forehead temperature to display its “oral equivalent.”
(according to the result of the clinical evaluation to get the offset value)
þ Blue LED Backlight
Blue LED backlight will be automatically turned on after measurement, and automatically
turned off after 2 seconds.
þ Enclosure Rating: IP22
þ Dimensions: 158.0 x 48.0 x 40.2 mm
þ Weight: 100 grams including battery
The device should not submerge into any liquids and expose it to direct moisture.
There is no gender and age limitation for using the infrared thermometer.
This is not an AP or APG product.
icon.
THD2FE Non-contact Forehead Thermometer
Functions
The thermometer has been designed for home use. It’s not meant to replace a visit to the
doctor. Please also remember to compare the measurement result to your regular body
temperature. Please consult with doctor if you have health concerns.
à Please see the “Use of the thermometer”
section to learn how to measure the body
temperature.
The surface mode shows the actual and unadjusted surface temperature which is different from
the body temperature. It can help you monitor if the object temperature is suitable for the baby
or patient, for example the baby’s milk.
à Please see the “Use of the thermometer” section to learn how to measure
the object
temperature.
If the thermometer detects a temperature ≧99.5°F (or 37.5°C) under forehead mode, three
short beep sound will follow one long beep sound to warn the user for the temperature is above
99.5°F (37.5°C).
There are total 25 sets of measurement condition for body temperature.
à When power on, press the “ON/MEM” button to see the temperature records with icon.
In “Power Off” mode, press and hold the “START” button, then press the “ON/MEM” button for
3 seconds, icon “°F” will be switched to icon “°C”. You can also use the same process to
change the LCD display from °C to °F.
NOTE:
Memory clear and Celsius/Fahrenheit switch are together. When you switch Celsius and
Fahrenheit, the Memory will be cleared.
The device setting with buzzer is on, you can set buzzer on/off under Mute mode.
When power on, press and hold the “ON/MEM” button for 3 seconds. The icon
will flash on
the LCD screen and then release the “ON/MEM” button to set MUTE. Thus you will not hear
beep sounds. You can also use the same process to turn off the Mute function.
NOTE: If keep pressing "ON/MEN" button for 2 seconds after
icon flashing, the device will
be power off WITHOUT setting Mute.
Use of the thermometer
Note: If there is any temperature difference between the places where the device is stored and where you are going to measure, subject and the device should
stay in the same room for at least 15 minutes before measurement.
1. Always make sure the probe lens is clean without any damage and the forehead is clean.
Warning: Choking from swallowing small parts and batteries by children or pets is possible, please keep small parts and batteries at places where
children and pets can’t reach.
2. Power on:
Press the “ON/MEM” button (see figure 1).
3. Measuring body temperature on the forehead:
The Non-contact Clinical Thermometer, Model THD2FE is an infrared thermometer intended
for the intermittent measurement of human body temperature in people of all ages.
At least 11 years old (5 years intensive reading experience), no maximum.
þ Temperature measurement range:
Forehead mode: 93.2~108°F (34~42.2°C)
Surface mode: -7.6~176°F (-22~80°C)
þ Operating temperature range: 50~104°F (10~40°C), 15%~85% RH
þ Storage temperature range: It should be stored at room temperature between -4~122°F
(-20~+50°C), RH≦85%
Transportation temperature shall be less than 158°F (70°C), RH≦95%
þ Atmospheric pressure: 800~1013 hPa
þ Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC),
IEC/EN60601-1(Safety) standards, ISO10993, RoHS.
þ Accuracy:
Forehead mode: ±0.4
o
F (0.2
°C ) within 95~107.6°F (35~42°C),
±0.5
o
F (0.3
°C ) for other range.
Surface mode: ±0.5°F (0.3°C) within 71.6~108°F (22~42.2°C),
others ±4% or ±4°F (2°C) whichever is greater.
þ Temperature indicator, memory and °F / °C switch function
þ Battery: AAA x 2 pcs
þ Battery life: around 3,000 continuous readings.
þ Expected Service Life: 4 years
þ This thermometer converts the forehead temperature to display its “oral equivalent.”
(according to the result of the clinical evaluation to get the offset value)
þ Blue LED Backlight
Blue LED backlight will be automatically turned on after measurement, and automatically
turned off after 2 seconds.
þ Enclosure Rating: IP22
þ Dimensions: 158.0 x 48.0 x 40.2 mm
þ Weight: 100 grams including battery
The device should not submerge into any liquids and expose it to direct moisture.
There is no gender and age limitation for using the infrared thermometer.
This is not an AP or APG product.
START
LENTILLE
SONDE
STAND-BY
(ON-MEM)
LCD
COUVERCLE
PILES
MARCHE
Français