MANUFACTURER’S DECLARATION-ELECTROMAGNETIC EMISSIONS
Intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user
should assure that it is used in such an environment.
EMISSIONS
TEST
COMPLIANCE ELECTROMAGNETIC ENVIRONMENT - guidance (for home healthcare environment)
RF emissions
CISPR 11
Group 1
This unit uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
This unit is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
Intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user
should assure that it is used in such an environment.
IMMUNITY TEST IEC 60601 TEST LEVEL COMPLIANCE LEVEL
ELECTROMAGNETIC ENVIRONMENT - guidance
(for home healthcare environment)
Electrostatic discharge
(ESD)
IEC 61000-4-2
Contact: ±8 kV
Air ±2 kV, ±4 kV, ±8
kV, ±15 kV
Contact: ±8 kV
Air ±2 kV, ±4 kV, ±8
kV, ±15 Kv
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%
Power frequency (50,
60 Hz) magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz and 60 Hz
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical home healthcare environment.
MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
Intended for use in the electromagnetic environment (for home healthcare specified below. The customer or the user
should assure that it is used in such an environment.
IMMUNITY TEST
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVIRONMENT - guidance (for home
healthcare environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7
GHz
80% AM at 1
kHz
10 V/m
80 MHz – 2,7
GHz
80% AM at 1
kHz
Recommended separation distance:
d = 1,2
Please read the instructions for use
Classification for water ingress and
particulate matter.
Manufacturer’s declaration-electromagnetic emissions
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
The THD2FE uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
The THD2FE is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Manufacturer’s declaration – electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
Electrostatic discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 Kv
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%
Power frequency(50, 60 Hz)
magnetic field IEC 61000-4-8
The THD2FE power frequency magnetic fields should be at levels
characteristic of a typical location in a typical home healthcare
environment.
Manufacturer’s declaration – electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
Recommended separation distance:
d = 1,2
d = 1,2 80MHz to 800 MHz
d = 2,3 800MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the THD2FE is used exceeds the
applicable RF compliance level above, the THD2FE should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the THD2FE.
Recommended separation distances between
portable and mobile RF communications equipment and the THD2FE
d = 1,2
Please read the instructions for use
Classification for water ingress and
particulate matter.
Manufacturer’s declaration-electromagnetic emissions
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
The THD2FE uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
The THD2FE is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Manufacturer’s declaration – electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
Electrostatic discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 Kv
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%
Power frequency(50, 60 Hz)
magnetic field IEC 61000-4-8
The THD2FE power frequency magnetic fields should be at levels
characteristic of a typical location in a typical home healthcare
environment.
Manufacturer’s declaration – electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
Recommended separation distance:
d = 1,2
d = 1,2 80MHz to 800 MHz
d = 2,3 800MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the THD2FE is used exceeds the
applicable RF compliance level above, the THD2FE should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the THD2FE.
Recommended separation distances between
portable and mobile RF communications equipment and the THD2FE
80 MHz to 800 MHz
d = 2,3
Please read the instructions for use
Classification for water ingress and
particulate matter.
Manufacturer’s declaration-electromagnetic emissions
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
The THD2FE uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
The THD2FE is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Manufacturer’s declaration – electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
Electrostatic discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 Kv
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%
Power frequency(50, 60 Hz)
magnetic field IEC 61000-4-8
The THD2FE power frequency magnetic fields should be at levels
characteristic of a typical location in a typical home healthcare
environment.
Manufacturer’s declaration – electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
Recommended separation distance:
d = 1,2
d = 1,2 80MHz to 800 MHz
d = 2,3 800MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the THD2FE is used exceeds the
applicable RF compliance level above, the THD2FE should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the THD2FE.
Recommended separation distances between
portable and mobile RF communications equipment and the THD2FE
800 MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
a) should be less than the compliance level in each frequency
range.
b) interference may occur in the vicinity of equipment marked
with the following symbol:
Please read the instructions for use
Classification for water ingress and
particulate matter.
Manufacturer’s declaration-electromagnetic emissions
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
The THD2FE uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
The THD2FE is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Manufacturer’s declaration – electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
Electrostatic discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 Kv
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%
Power frequency(50, 60 Hz)
magnetic field IEC 61000-4-8
The THD2FE power frequency magnetic fields should be at levels
characteristic of a typical location in a typical home healthcare
environment.
Manufacturer’s declaration – electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Electromagnetic environment – guidance
(for home healthcare environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
Recommended separation distance:
d = 1,2
d = 1,2 80MHz to 800 MHz
d = 2,3 800MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment d ue to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the THD2FE is used exceeds the
applicable RF compliance level above, the THD2FE should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the THD2FE.
Recommended separation distances between
portable and mobile RF communications equipment and the THD2FE
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the thermometer is used exceeds the applicable
RF compliance level above, it should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the unit.
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT
AND THE THERMOMETER
Intended for use in an electromagnetic environment (for home healthcare) in which radiated RF disturbances are
controlled. The customer or the user can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the unit as recommended below,
according to the maximum output power of the communications equipment.