6 www.mastercool.com
RATED MAXIMUM OUTPUT
POWER OF TRANSMITTER
W
SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER
m
150 kHz to 80 MHz
d = 1,2
Symbol Descriptions
Caution
Please read the instructions for use
Battery Recycling
BF type applied part
Paper Recycling
Stand-by
IP22
Classification for water ingress and
particulate matter.
Manufacturers declaration-electromagnetic emissions
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the THD2FE
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment guidance
(for home healthcare environment)
RF emissions
CISPR 11
Group 1
The THD2FE uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The THD2FE is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Manufacturers declaration electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
(for home healthcare environment)
Electrostatic discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Ai2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Ai2 kV,±4 kV,±8
kV,±15 Kv
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%
Power frequency(50, 60 Hz)
magnetic field IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz and 60 Hz
The THD2FE power frequency magnetic fields should be at levels
characteristic of a typical location in a typical home healthcare
environment.
Manufacturers declaration electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment guidance
(for home healthcare environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz 2,7
GHz
80 % AM at 1
kHz
Recommended separation distance:
d = 1,2
d = 1,2 80MHz to 800 MHz
d = 2,3 800MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the THD2FE is used exceeds the
applicable RF compliance level above, the THD2FE should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the THD2FE.
Recommended separation distances between
portable and mobile RF communications equipment and the THD2FE
80 MHz to 800 MHz
d = 1,2
Symbol Descriptions
Caution
Please read the instructions for use
Battery Recycling
BF type applied part
Paper Recycling
Stand-by
IP22
Classification for water ingress and
particulate matter.
Manufacturers declaration-electromagnetic emissions
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the THD2FE
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment guidance
(for home healthcare environment)
RF emissions
CISPR 11
Group 1
The THD2FE uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The THD2FE is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Manufacturers declaration electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
(for home healthcare environment)
Electrostatic discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Ai2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Ai2 kV,±4 kV,±8
kV,±15 Kv
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%
Power frequency(50, 60 Hz)
magnetic field IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz and 60 Hz
The THD2FE power frequency magnetic fields should be at levels
characteristic of a typical location in a typical home healthcare
environment.
Manufacturers declaration electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment guidance
(for home healthcare environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz 2,7
GHz
80 % AM at 1
kHz
Recommended separation distance:
d = 1,2
d = 1,2 80MHz to 800 MHz
d = 2,3 800MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the THD2FE is used exceeds the
applicable RF compliance level above, the THD2FE should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the THD2FE.
Recommended separation distances between
portable and mobile RF communications equipment and the THD2FE
800 MHz to 2,7 GHz
d = 2,3
Symbol Descriptions
Caution
Please read the instructions for use
Battery Recycling
BF type applied part
Paper Recycling
Stand-by
IP22
Classification for water ingress and
particulate matter.
Manufacturers declaration-electromagnetic emissions
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the THD2FE
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment guidance
(for home healthcare environment)
RF emissions
CISPR 11
Group 1
The THD2FE uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The THD2FE is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Manufacturers declaration electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
(for home healthcare environment)
Electrostatic discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Ai2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Ai2 kV,±4 kV,±8
kV,±15 Kv
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%
Power frequency(50, 60 Hz)
magnetic field IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz and 60 Hz
The THD2FE power frequency magnetic fields should be at levels
characteristic of a typical location in a typical home healthcare
environment.
Manufacturers declaration electromagnetic immunity
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the THD2FE
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment guidance
(for home healthcare environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz 2,7
GHz
80 % AM at 1
kHz
Recommended separation distance:
d = 1,2
d = 1,2 80MHz to 800 MHz
d = 2,3 800MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the THD2FE is used exceeds the
applicable RF compliance level above, the THD2FE should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the THD2FE.
Recommended separation distances between
portable and mobile RF communications equipment and the THD2FE
0,01 N/A 0,12 0,23
0,1 N/A 0,38 0,73
1 N/A 1,2 2,3
10 N/A 3,8 7,3
100 N/A 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
MANUFACTURER’S DECLARATION-ELECTROMAGNETIC IMMUNITY TEST SPECIFICATIONS FOR ENCLOSURE PORT
IMMUNITY TO RF WIRELESS COMMUNICATIONS EQUIPMENT
Intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user
should assure that it is used in such an environment.
TEST
FREQUENCY
(MHz)
BAND a)
(MHz)
SERVICE
a)
MODULATION
b)
MAXIMUM
POWER
(W)
DISTANCE
(m)
IMMUNITY
TEST
LEVEL
(V/m)
COMPLIANCE
LEVEL (V/m)
(for home health-
care)
385
380 -
390
TETRA 400
Pulse modula-
tion b) 18 Hz
1,8 0,3 27 27
450
430 -
470
GMRS 460,
FRS 460
FM c)
±5 kHz deviation
1 kHz sine
2 0,3 28 28
710
704 -
787
LTE Band 13,17
Pulse
modulation b)
217 Hz
0,2 0,3 9 9745
780
810
800 -
960
GSM 800/900,
TETRA 800, iDEN
820, CDMA 850, LTE
Band 5
Pulse
modulation b)
18 Hz
2 0,3 28 28
870
930
1720
1700 -
1990
GSM 1800; CDMA
1900; GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse modula-
tion b) 217 Hz
2 0,3 28 28
1845
1970
2450
2400 -
2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation b)
217 Hz
2 0,3 28 28
5240
5100 -
5800
WLAN 802.11
a/n
Pulse
modulation b)
217 Hz
0,2 0,3 9 95500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent
actual modulation, it would be worst case.