Hydrogen Generator NM Series 2 User’s Manual
9
Installing the deionizer bag
The triangle deionizer bag has been designed for a higher water purifying capacity. It is rec-
ommendable to use this bag for new generators, in the first 4 to 6 months of operation.
After this time you can use the standard deionizer bag (see “spare parts”).
Fig. 8 Triangle deionizer bag Fig. 9 Standard deionizer bag
After having filled the tank with water, the triangle deionizer bag (supplied) must be placed in
the tank. Inspect the bag thoroughly for holes or tears, indicated by loose deionizer beads on
the outer surface. If the bag is damaged in any way, discard and replace it with a new one.
Only use original parts (see Spare Parts). Wash the deionizer bag in deionized water before
proceeding.
Insert the free end of the “T” fastener through one of the holes of the holder, until it is se-
curely fastened (Fig. 10). Insert the bag in the opening of the tank (Fig 11).
Fig. 10 Fig. 11