Example: When the load address is 1, set the constant current value IFIX to 2.3
Table 4.8.7 Table .2 to know the current value IFIX register address is 0x0A01, a 2-word floating-point number
Send request: 0x01 0x10 0x0A 0x01 0x00 0x02 0x04 0x40 0x13 0x33 0x33 0xFC 0x23
Get the normal reply 0x01 0x10 0x0A 0x01 0x00 0x02 0x13 0xD0
4.8.7 Coil and register address assignment
Table 1 coil bit definition:
Remote control status bit: When 1, the front button
panel is invalid.
Local Prohibit bit: 1, prohibit the use of "Shift + 7"
key to regain control of the front panel.
Trigger marker bit, software can be set to complete
a trigger.
Enter the state, 1 for the input ON, 0 for the input
OFF
Tracking status, 1 for voltage tracking, 0 for current
tracking
1 is the input status memory
1 is multi-machine, 0 is single work mode
1 is the automatic test mode of operation
1 Awaiting trigger for automatic test mode of
operation
1 for automatic test passed, 0 failed
1 is the overcurrent flag
1 is a fixed parameter failure flag
1 is the EPPROM error flag
1 is the calibration data error flag