ERROR = P. LOW
Open the gas bottle
Verify the set pressure on the regulator
(50-60 psi, 3.4-4.1 bar or 340-410 kPa)
Verify the gas line
Verify the reading of the pressure
transducer «STATUSPRESSURE»
ERROR = P. HIGH
Verify the reading of the pressure
transducer «STATUSPRESSURE»
You may have a solenoid valve
malfunction. Restart the generator.
Please contact MDG Service if this problem
persists.
ERROR = HEATER
This is a heating timeout, due to a
cartridge heater problem.
Restart the generator and check the
heating status «STATUSSTATE = xx%
HEAT». If the heating value is not
progressing, there may be various issues.
Check the fuses on the heater board
Contact MDG Service.
ERROR = T. HIGH
Restart the generator. This is generally
due to an electronic problem.
Please contact MDG Service if this problem
persists.
ERROR = T. SAF
Restart the generator. This is generally
due to an electronic or sensor problem.
Please contact MDG Service if this problem
persists.