page 16
The display will show:
RESET When the load limit is reached, the output
signal triggers and the RESET key must be
pressed to clear the line before starting the
next test.
CONTINUOUS The output signal will be activated every
time the load limit is reached and will
remain on whilst the load exceeds the set
limit.
PULSE The output signal will be activated
momentarily when the load limit is
reached. The RESET key must be pressed
before starting the next test.
Select the desired function and press the ENTER key.
The display will show SET and a default load limit at which
the output signal will trigger. To set the required load limit use
UP and DOWN keys to adjust the value and the ENTER key
to confirm the selection.
The display will show STATE:
HIGH Will set PLC signal high at Tornado's alarm.
LOW Will set PLC signal low at Tornado's alarm.
Select the desired function and press the ENTER key.
The display will revert back to PLC sub-menu 1 and PLC ON
will now be displayed.
Press ESC key to return to the main menu page 1.
AT LIMITS sub-menu 1
AT LIMITS sub-menu 2
AT ALARM sub-menu 1