Alarms and Troubleshooting
46
No Drop Sensor
Middle
The drop detection
function is enabled but
the drop sensor is not
installed during the
infusion.
Install the drop sensor, or
stop infusion and disable
the drop detection
function.
No Drop
High
No drop is detected
during the infusion.
Tap Stop to eliminate the
alarm. Check the
installation of the drop
sensor and the status of
the infusion tube.
Too Many Drops
High
The drop rate is too high
during the infusion.
Tap Stop to eliminate the
alarm. Check the
installation of the drop
sensor.
Too Few Drops
High
The drop rate is too low
during the infusion.
Tap Stop to eliminate the
alarm. Check the
installation of the drop
sensor.
Drop Chamber
Overflow
High
The drip chamber
overflows during the
infusion.
Tap Stop to eliminate the
alarm. Check the
installation of the drop
sensor.