Alarms and Troubleshooting
45
1. The syringe line is
occluded.
2. A low occlusion level
is set for high-viscosity
solution infusion.
3. The system will auto
reduce the volume when
an occlusion occurs.
Tap Stop to eliminate the
alarm. Identify and
remove the occlusion
cause and resume the
infusion.
Tap Cancel to exit the
standby state.
No key operation is
performed in 2 min after
the syringe is installed.
Tap any key to eliminate
the alarm.
The communication is
interrupted in the relay
process or any external
interference causes relay
sequence failure.
Tap Stop to eliminate the
alarm. Enable the local
relay again.
9.3 Infusion Pump Faults and Troubleshooting
The following table lists the alarm message dedicated for the infusion pump and the alarm level, alarm cause,
and troubleshooting.
Tap Stop to eliminate the
alarm. Identify and
remove the occlusion
cause and resume the
infusion.
1. There is air bubble
inside the infusion tube.
2. Damaged infusion set
is installed on the air
bubble sensor.
Tap Stop to eliminate the
alarm. Check that the part
of the infusion set
installed on the air bubble
sensor does not deform,
and remove the air
bubble.