REPLACEMENT OF THE NEBULIZER
The nebulizer must be replaced after a long period of inactivity, or when it is obstructed by dry
medication, dust, etc.
Your nebulizer should last from 6 months to 1 year with general use.
Use original nebulizers only.
REPLACEMENT OF THE FILTER
The MEDEL PRO Porex filter must be replaced approximately after 500 hours of use or each year;
control the filter periodically (10-12 aerosol treatments ); when it turns grey or brown or wet,
replace it; open the filter holder (pic. A/5) and replace the filter with a new one. Don’t try to clean
the filter and to reuse.
Use original filters only.
Do not use the unit without filter.
POSSIBLE PROBLEMS AND HOW TO SOLVE THEM
If the device does not switch on:
1. make sure the plug is firmly fitted to the wall socket;
2. the thermal protector might have been released. The thermal protector may be released for
the following reasons:
- the device has been in operation for a too long time;
- the device was used near heaters or in hot environments;
- the air vent openings (pic. A/7) are obstructed and do not allow for the motor ventilation.
In this case, set the ON/OFF switch to “O” and let it rest for at least 60 minutes.
If the device does not nebulize, or only ver
y weakly, check the following
points:
1. make sure that the ends of the air tube are fitted tightly onto the main unit and onto the
nebulizer;
2. check whether the nebulizer is empty from medication or full of an excessive medication
quantity (MAX 15 ml);
3. check whether the nebulizer nozzle is obstructed;
4. check whether the air conduction cone is inserted (pic. E).
Should the device not start working properly again, address immediately to
an authorized technical service centre or to MEDEL CUSTOMER SERVICE.