MAINTENANCE AND REPAIRS
In case of failure, address to qualified personnel authorized by Medel S.p.A.. Never open the
device. There are no user-serviceable parts within. The compressor requires no lubrication or
maintenance.
TECHNICAL SPECIFICATIONS
Model: MEDEL PRO
Electricity supply: see rating label on the device bottom
Nebulizer air flow: 3,3 l/min
Compressor air flow: 7,50 l/min
Compressor max. pressure: 250 kPa (2,50 bar)
Weight: 1,91 kg
Size: (H) 106 x (L) 198 x (P) 233 mm
Noise Level: 52 dBA
Nebulization rate: ca. 0,5 ml/min
Particle size: MMAD 1,8 µm (NaCl) – API AEROSIZER;
MMD 4,17 µm – INSITEC MALVERN
MEDELJET PRO
nebulizer capacity: 15 ml (MAX)
Class II Device (Double Safety Insulation)
Type B Device (device with specific protection against electrical hazards)
Device for continuous use.
Device not protected against sprinkles
Device not suitable for use in presence of anaesthetic substances inflammable with air, oxygen or
nitrogen protoxide.
Device not suitable for use in anaesthesia or pulmonary ventilation systems.
Technical modifications reserved!
Reference to Standards: Electric Safety Standards CEI EN 60601-1
Electromagnetic Compatibility according to CEI EN 60601-1-2
Class II Medical Device according to 93/42/EEC
“Medical Device” Directive.
AMBIENT CONDITIONS
Storing Conditions
Temperature: MIN -25°C - MAX 70°C
Humidity: MIN 10%RH - MAX 95%RH
Operating Conditions
Temperature: MIN 10°C - MAX 40°C
Humidity: MIN 10%RH - MAX 95%RH
Altitude: from 0 to 2000m above sea level