Error codes
The measurement timeout is over, the
measurement had to be aborted (the patient
was moving)
The measurement was stopped by pressing
a button. The display differs from others:
"OFF" on the LCD.
The batteries exhausted during
measurement.
The batteries were replaced during
measurement.
E8: pressure limit exceeded
The pressure in the pneumatic system
exceeded the preset pressure limit
E9: temporary disturbance
External electric signal - static discharge -
disturbed the operation of the device.
E31: cuff missing or loose
There was no cuff connected to the device
or the cuff was too loose on the patient's
arm.
The cuff is clogged or the rubber tube is
broken.
E33: cuff leaking or loose
There is a hole in the cuff or it is rather
loose on the patient's arm.
The patient did not wear the cuff.
The device could not measure due to a
hardware error.
The device does not start automatic
measurement until the next programming.
Meditech product warranty information
RECORDER WARRANTY. The main recorder unit will be free from defects in
materials and workmanship under normal use and service for a period of two (2)
years from the date of receipt. This warranty covers the recorder unit only. This
warranty does not cover any accessories that might come with the recorder unit.
ACCESSORIES WARRANTY. The non-disposable accessories delivered with the
recorder unit will be free from defects in materials and workmanship under normal
use and service for a period of one (1) year from the date of receipt. This warranty
does not cover disposable accessories, packaging materials, accumulators and
batteries, cuffs, or any of their components.
CUFF WARRANTY. The cuff(s) - if delivered with a recorder unit - will be free from
defects in materials and workmanship under normal use and service for a period of
six (6) months from the date of receipt. This warranty covers the cuff(s) delivered
with a recorder unit exclusively.