BP5BB5LR_UMEN_V05-20170703 26
EMC information
Medical electrical equipment should be used with precautions according to EMC,
and must be installed according to the EMC notices disclosed in this manual as
mobile RF transceivers could adversely affect it.
Electromagnetic emission
Meditech ambulatory blood pressure monitors are suitable for use in the specified
electromagnetic environment. The purchaser or user of the device should assure
that they are used in an electromagnetic environment as described below.
Electromagnetic environment
Radiated and
conducted RF emission
CISPR 11
ABPM-05/BlueBP-05 uses RF energy only for their
internal function. Therefore, the emission is very
low and it is not likely to cause any interference in
nearby electronic equipment. Except for Bluetooth
communication usage. (The Bluetooth
radiofrequency communication is carried out at the
2.4GHz frequency; class2, Bluetooth 2.0+EDR)
Radiated and
conducted RF emission
CISPR 11
ABPM-05/BlueBP-05 is suitable for use in
domestic establishments and in establishments
directly connected to the low voltage power supply
network which supplies buildings used for
domestic use.
Harmonic emission
IEC61000-3-2
Voltage fluctuations/
Flickers IEC61000-3-3
Electromagnetic immunity
Meditech ambulatory blood pressure monitors are suitable for use in the specified
electromagnetic environment. The purchaser or user of Meditech products should
assure that they are used in an electromagnetic environment as described below.
Electromagnetic environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
Floors are wood, concrete or
ceramic tile, or floors are
covered with synthetic
material and the relative
humidity is at least 30
percent.
Electrical fast
transient/burst
IEC 61000-4-4
+ 2KV for power
supply
+ 1KV input/output
lines
Mains power quality is that of
a typical commercial and/or
hospital environment.
+ 1KV differential
mode
+ 2KV common
mode
Mains power quality is that of
a typical commercial and/or
hospital environment.