Start button functions
- to start a manual blood pressure measurement (press shortly)
Typical causes for this use: dizziness, pain (angina pectoris or headache),
palpitation.
- to switch the device off (press and hold for more than 10 seconds)
Press and hold the Start button until 2 horizontal segments appear on the
LCD ( - - ).
- to switch the device on (press and hold for more than 3 seconds)
- to check the LCD
Press and hold the start button to light up all segments of the LCD to check
if they all work correctly.
- to check battery voltage (press and hold for more than 5 seconds but less
than 10 seconds)
The voltage for fully charged accumulators should be over 2,5V (2_50 on the
LCD).
Event button functions
- to set a patient event marker (press shortly)
Typical cause for this use is taking medicine. The patient should be
instructed to record the reason for setting an event marker in a diary.
Day/night button functions
- Marking the time of sleeping and awakening
If the day/night shift function is disabled during programming, the patient
can press the day/night button to mark the time of sleeping (in the evening)
and awakening (in the morning).
- Manual day/night shift (only in the 2 hour period before the prescheduled
shift)
If this function is enabled during the programming, the patient can manually
shift the measurement frequency period (day or night) by pressing the
day/night button.
Bluetooth functions – BlueBP-05 only
- Switching to Bluetooth discoverable mode
to start communication
You can initiate the Bluetooth mode by
pressing the Start and the Event buttons
simultaneously for longer than 3 seconds
then release them, until you can see this on
the LCD screen: PC --.
- Switching off Bluetooth discoverable mode
You can switch off the Bluetooth mode by
pressing the Start and the Event buttons
simultaneously. The discoverable mode is
automatically ceased after reading out or
programming or if the connection fails to be
established in one minute.