DSD™-201
Endoscope
Reprocessor
User
Manual
OPERAT
Running
the
Disinfection
Process
1.
Place
the
floating
lid
on
the
basin.
Verify
the
endoscope
or
hook-up
does
not
pro-
trude
from
the
basin
or
contact
the
floating
basin
lid.
2.
Close
the
reprocessor
lid.
3.
Press
the
STATION
SELECT
button
to
choose
station
A
or
station
B.
4.
Press
the
ID
DATA
button,
then
enter
the
ID
Data
(if
required
for
printed
log).
-
Enter
up
to
ten
digits
for
the
endoscope
serial
number,
then
press
the
ENTER
button.
ION
-
Enter
up
to
ten
digits
for
the
operator
ID
number,
then
press
the
ENTER
button.
-
Enter
up
to
ten
digits
for
the
patient
ID
number,
then
press
the
ENTER
button.
-
Enter
up
to
ten
digits
for
the
physician
ID
number,
then
press
the
ENTER
button.
5.
Select
the
desired
disinfection
program
on
the
reprocessor
control
panel.
-
Select
0
for
the
default
program.
-
Select
1-9
for
a
custom
program.
6.
Press
the
START
button.
-
The
lid
locks
engage
(optional)
and
the
disinfection
program
starts.
-
Ifthe
leak
tester
option
is
enabled,
there
is
a
40
second
delay
at
the
beginning
of
the
cycle.
7.
During
the
detergent
flush
cycle,
verify
the
endoscope
connections
and
plugs
are
properly
connected.
-
Verify
fluid
flows
through
the
channels.
-
Verify
fluid
flows
from
the
distal
end
of
the
endoscope.
-
Verify
there
are
no
leaks
at
the
channel
fittings
and
adaptors.
8.
Indicators
on
the
control
panel
display
status
information
while
the
reprocessor
pro-
cesses
the
endoscope.
A
reg
To
interrupt
the
process
at
any
time,
or
to
clear
errors,
refer
to
the
Process
Interruption
procedure
in
Chapter
5
of
this
manual.
©
2008,
Medivators
Reprocessing
Systems
129