REPROCESSING
SYSTEMS
Electrical
Supply
Domestic
reprocessors
are
supplied
with
a
hospital
grade,
grounding
plug
that
can
be
connected
to
any
standard
110-130VAC
outlet.
International
reprocessors
are
supplied
with
a
standard
ground-
ing
European
plug
that
connects
to
220-240VAC
outlets.
We
recommend
surge
protectors
be
used
for
protection
against
power
spikes
and
surges.
The
reprocessor
must
be
plugged
into
or
connected
to
a
fused
branched
circuit.
Caution!
The
reprocessor
does
not
have
an
on/off
switch.
Be
sure
the
reprocessor
is
positioned
so
that
the
power
cord
or
main
circuit
breaker
is
accessible
at
all
times.
Warning!
The
reprocessor
must
be
protectively
grounded.
ας
Air
Supply
Compressed
air
is
required
(30
-
60
psi/2.1
-
4.1
bar).
If
you
did
not
purchase
the
air
compressor
option
with
the
reprocessor,
be
sure
to
make
the
compressed
air
connection
to
the
labeled
fitting
on
the
side
of
the
cabinet.
Warning!
This
air
must
be
supplied
from
an
oil
free
air
source.