174
MM03-0047
Revision
D
REPROCESSING
SYSTEMS
"Res
T
Low”
is
displayed.
Incorrect
reservoir
controller
temperature
..........
Adjust
to
correct
temperature.
Incorrect
program
temperature
setting................
Use
Diagnostics
74
to
enter
value.
Thermistor
defective
........................
n
Repair
or
replace
the
component.
Heater
turned
off
or
defective
.............................
Turn
heater
on
or
replace.
Check
harness
connection.
“Sheath
Fail”
is
displayed
(for
optional
leak
tester
only).
Large
leak
detected
at
beginning
of
cycle...........
Leak
in
scope.
Allow
the
scope
to
reprocess,
then
remove
scope
and
send
for
repair.
Leak
between
connectors.
Press
STOP
to
cancel
the
cycle
and
open
the
lid
locks.
Check
connec-
tion
and
reprocess
scope.
Small
leak
detected
during
cycle
........................
Leak
in
scope.
Allow
the
scope
to
reprocess,
then
remove
and
manually
leak
test
the
scope.
Leak
test
adapter
not
connected..........................
Press
STOP
to
cancel
the
cycle
and
open
the
lid
locks.
Connect
correct
leak
test
adapter
and
reprocess
scope.
Leak
tester
harness
disconnected/defective
........
Check
the
leak
tester
harness
connection.
Replace
a
defective
harness.
Leak
tester
board
defective..................................
Replace
the
leak
tester
board.
Leak
tester
assembly
defective
...........................
Replace
the
leak
tester
assembly.
"Shth
Sen
Err”
is
displayed
during
initial
start
(for
optional
leak
tester
only).
Scope
is
pressurized
during
startup
....................
Press
STOP
to
cancel
the
cycle
and
open
the
lid
locks.
Disconnect
the
leak
tester
hook-up
to
release
the
pressure.
Reconnect
the
hook-up.
Leak
tester
enabled,
but
option
not
present
........
Use
Diagnostic
45
to
disable
leak
tester
option.
Leak
tester
board
defective
.................................
Replace
the
leak
tester
board.
Leak
tester
defective
......................
Replace
the
leak
tester.