DSD-201™
Endoscope
Disinfector
Service
Manual
TROUBLESHOOTING
Disinfectant
filter
is
disconnected.......................
Reconnect
the
disinfectant
filter.
Scope
channel
is
blocked....................................
Remove
scope
and
send
for
repair.
Sensor
harness
disconnected/defective...............
Check
the
sensor
harness
connection.
Replace
a
defective
harness.
Sensor
board
defective........................................
Replace
the
sensor
board.
Sensor
defective
iii
Verify
proper
fluid
flow.
Replace
the
sensor.
“No
Fluid
Flow”
is
displayed
during
flush
or
rinse
phase.
Scope
to
basin
connection
disconnected.............
Reconnect
and
restart
the
cycle.
WaterpressuretooloW........................................
Pressure
must
be
between
35-40psi
(2.4
-
2.75bar).
Prefilter/internal
filter
clogged............................
Change
the
filter.
Air
lock
in
the
water
filter...................................
Open
the
bleeder
valve
on
the
filter
to
purge
air.
Waterfilterisclogged.......................................
Check
the
water
pressure
at
inlet
and
outlet
|
regulators.
Lower
than
normal
pressure
may
y
indicate
a
clogged
filter.
Replace
the
filter.
Air
pressure
switch
harness
disconnected...........
Check
the
harness
for
proper
connection.
Also,
check
for
damage,
loose
wires.
Sensor
harness
disconnected/defective
...............
Check
the
sensor
harness
connection.
Replace
a
defective
harness.
Sensor
board
defective........................................
Replace
the
sensor
board.
Sensor
defective
in
Verify
proper
fluid
flow.
Replace
the
sensor.
"Res
T
High”
is
displayed.
Incorrect
reservoir
controller
temperature
..........
Adjust
to
correct
temperature.
Incorrect
program
temperature
setting................
Use
Diagnostics
74
to
enter
value.
Thermistor
defective
...........................
iii
Repair
or
replace
the
component.
©