-30-
of 10,000 measurements or three years.
10. It is recommended the cuff should be disinfected 2 times every week if needed
(For example, in hospital or in clinic.). Wipe the inner side (the side that contacts
skin) of the cuff with a soft cloth squeezed after being moistened with Ethyl alcohol
(75-90%), then dry the cuff by airing.
11. The cuff integrity is maintained after 1,000 open-close cycles of the closure.
12. The cuff will maintain performance characteristics for a minimum of 10,000 inflations.
13. Not servicing/maintenance while the monitor is in use.
14. The monitor requires 6 hours to warm from the minimum storage temperature
between uses until the monitor is ready for its INTENDED USE when the ambient
temperature is 20 °C.
15. The monitor requires 6 hours to cool from the maximum storage temperature
between uses until the monitor is ready for its INTENDED USE when the ambient
temperature is 20 °C.