-35-
Guidance and manufacturer’s declaration - Proximity fields from RF wireless
communications equipment
The [MDS3001/MDS3001PLUS/MDS3001U] is intended for use in the electromagnetic environment
specified below. The customer or the user of the [MDS3001/MDS3001PLUS/MDS3001U] should assure
that it is used in such an environment.
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385 380-390 Pulse modulation 18Hz, 27V/m
450 430-470 FM, ±5kHz deviation, 1kHz sine, 28V/m
710
704-787 Pulse modulation 217Hz, 9V/m745
780