Insulation Multimeters
User Guide
SAFETY WARNINGS
Safety Warnings and Precautions must be read and
understood before the instrument is used. They
must be
observed during use.
The circuit under test must be de-energised and isolated
before connections are made except for voltage measurement.
Circuit connections must not be touched during a test.
After insulation tests, capacitive circuits must be allowed to
discharge
before disconnecting the test leads.
The Live Circuit Warning and Automatic Discharge are additional
safety features and
should not be regarded as a substitute for
normal safe working practice.
Replacement fuses must be of the correct type and rating.
Failure to fit the correctly rated fuse
will result in damage to the
instrument in the event of an overload.
Test leads, including crocodile clips, must be in good order,
clean and have no broken or cracked insulation.
Ensure that hands remain behind guards of probes/clips when
testing.
U.K. Safety Authorities recommend the use of fused test leads
when measuring voltage on high energy systems.
NOTE
THE INSTRUMENTS MUST ONLY BE USED BY
SUITABLY TRAINED AND COMPETENT PERSONS
Users of this equipment and/or their employers are reminded that
Health and Safety Legislation require them to carry out valid risk
assessments of all electrical work so as to identify potential
sources of electrical danger and risk of electrical injury such as
from inadvertent short circuits. Where the assessments show that
the risk is significant then the use of fused test leads constructed
in accordance with the HSE guidance note GS38 ‘Electrical Test
Equipment for use by Electricians’ should be used.
Symbols used on the instruments are:
Caution, risk of electric shock.
Caution, refer to User Guide.
Equipment protected throughout by
Double Insulation (Class II).
Equipment complies with current EU Directives.
Equipment must not be connected
to installations >500V.
>500V
Question and Answer IconNeed help?

Do you have a question about the Megger BMM2000 Series and is the answer not in the manual?

Megger BMM2000 Series Specifications

General IconGeneral
TypeDigital Multimeter
Continuity TestYes
Diode TestYes
Safety CategoryCAT III 1000 V, CAT IV 600 V
Voltage Measurement Range (DC)0.1mV to 1000V
Voltage Measurement Range (AC)0.1mV to 750V
Current Measurement Range (DC)0.1μA to 10A
Current Measurement Range (AC)0.1μA to 10A
Resistance Measurement Range0.1Ω to 50MΩ
DisplayLCD

Summary

Safety Warnings and Precautions

General Safety Precautions

Essential safety rules for instrument operation, including circuit isolation, lead handling, and fuse replacement.

User Competency and Risk Assessment

Emphasizes that instruments must only be used by trained personnel and highlights the need for risk assessments.

Instrument Operation Guide

Pre-Test Safety Inhibitions

Describes automatic inhibitions for external voltages on insulation and other ranges.

Live Circuit Warning and Voltage Testing

Details warnings for live circuits and precautions for high energy voltage testing.

Insulation Test Procedure

Step-by-step guide for performing insulation resistance tests, including connections and discharge.

Insulation Test Procedures and Best Practices

Guidance on accurate high-value measurements, lead care, and leakage current display.

Continuity and Resistance Testing

Continuity Testing (Ω)

Procedure for continuity tests, including activation, reading hold, and suitability.

Test Activation and Lead Resistance Zeroing

How continuity tests activate, hold readings, and how to zero test lead resistance.

Voltage and Millivolt Measurements

Voltage Measurement (V)

Procedure for measuring AC/DC voltage and displaying frequency.

Millivolt Measurement (mV)

Procedure for measuring AC/DC millivolts and selecting modes.

Capacitance and Milliamps Testing

Milliamps Tests (mA) (Not BMM2000ESD)

Procedure for mA measurement, including activation and mode switching.

Megger SP1 Switched Probe and Maintenance

SP1 Probe Safety and Features

Details the safety requirements, double insulation, and fuse protection of the SP1 probe.

Safety Compliance and Standards

Repair and Warranty Information