7 Sterilization
32
2. Confirm the security query with YES.
3. After a short time, you can open the door by pressing
the symbol. The display shows a warning;
the log records the sterilization asNOT
successful.
Program abort after the start of drying
Should you abort a program after drying has started, the sterilization is having been completed successful-
ly. The steam sterilizer issues a malfunction message. You then need to expect insufficient drying, espe-
cially in the case of wrapped sterile material and a full load. Sterile storage requires sufficient drying. To
ensure this, please allow programs with wrapped sterile material to continue to the end of the drying phase
as far as is possible. Unwrapped instruments sterilized in a Quick-Program dry from their own warmth after
being removed.
Proceed as follows to abort the program during drying:
1. Press STOP on the action bar.