12 Pause times
68
Emptying the double jacket
You have the option of draining the water in the double jacket steam generator easily via program Drain. In
order to do so, the steam sterilizer is heated once, building up pressure in the double jacket so that the wa-
ter can be drained fully from the double jacket steam generator.
1. Switch on the steam sterilizer at the power switch.
2. Working in menu Programs & Tests select
program Drain and press START.
3. Following notificationDraining successful
switch off the steam sterilizer, so that water is not
fed into the double jacket.
Transport
CAUTION
Danger of injury from incorrect carrying.
Lifting and carrying too heavy a load can result in spinal injury. Failure to comply with
these provisions can result in crushing.
n The device should always be carried by two people.
n Use the correct carrying straps to carry the device.
Transport within the practice
Comply with the following provisions during transport within a room or the practice:
u Decommission the steam sterilizer, see Decommissioning [}page67].
u Disconnect the connection hoses connected on the rear of the device.
u Should you wish to leave the mounts and trays or cassettes in the chamber during transport, protect
the surface of the round blank. To do so, place some foam or bubble wrap between the round blank
and mount.
u Close the steam sterilizer door before moving it.
Transport over long-distance / dispatch
For transport over longer distances, during the danger of frost and / for despatch or during maintenance,
an authorized person must prepare the steam sterilizer according to instructions and empty the double-
jacket steam generator entirely.
Recommissioning after relocation
When recommissioning after a move, proceed as with the first commissioning; see the technical manual.