After switching on the RQflex 10 a self-diagnosis is performed for about 1
second. Ensure that all functions, especially the multifunctional digit display,
are shown correctly.
Time 1
LOW BATTERY change of batteries 2
Date 3
Symbols for measurement procedure 4
Methods 5
Multifunctional display 6
Instrument, tests, and bar-code strips are tuned to each other. Ensure that the
reference number (code) for all three components matches.
Technical data
D i m e n s i o n s : 19 x 8 x 2 cm
We i g h t : 275 g
M e m o r y : 5 methods, 50 measur ements results
External output: Ye s
Light source: 4 LEDs (5 70 / 657 ± 10 nm), double optics
Power source: 4 AAA batteries (> 1000 measurements)
D i s p l a y : L C D
System diagnosis: Ye s
Measurement range: 4 - 90% rel. remission
Reflection area: 4 x 6 mm
R e s o l u t i o n : 0 .1% rel. remission
Photom. accuracy: 0 .5% rel. remission
(measurement to measurement, instrument to instrument)
Operating temperature: 5 - 40°C for ideal measurements
Operating humidity: Below 90 % for ideal measurements
|