MAINTENANCE
Non-approved disinfectants:
Organic solvents
Ammonia-based detergent
Abrasive cleaning agents
100% alcohol, acetone, chlorine etc.
Disinfection – cu:
Can be disinfected with medical-grade alcohol 70%.
After disinfection, the cu cover should be left air dry.
If correctly used and maintained the device will have a minimum
service life of 5 years. It is possible to store the device in suitable
conditions for a maximum of 5 years. While in storage the battery
needs to be recharged once per three months to avoid accidentally
over-discharging the batteries. When using the device after storage,
we recommend subjecting the device to a thorough maintenance
check.
The device should be inspected by a qualified service engineer at
least every 12 months for the following safety checks:
Any mechanical or functional damage on the device and
accessories,
Performance of device in accordance to instructions for use,
The warning label’s legibility, and
The battery cycle count.
6.4 PRODUCT LIFE AND STORAGE