41
MAINTENANCE
Non-approved disinfectants:
• Organic solvents
• Ammonia-based detergent
• Abrasive cleaning agents
• 100% alcohol, acetone, chlorine etc.
Disinfection – cu:
• Can be disinfected with medical-grade alcohol 70%.
• After disinfection, the cu cover should be left air dry.
If correctly used and maintained the device will have a minimum
service life of 5 years. It is possible to store the device in suitable
conditions for a maximum of 5 years. While in storage the battery
needs to be recharged once per three months to avoid accidentally
over-discharging the batteries. When using the device after storage,
we recommend subjecting the device to a thorough maintenance
check.
The device should be inspected by a qualified service engineer at
least every 12 months for the following safety checks:
• Any mechanical or functional damage on the device and
accessories,
• Performance of device in accordance to instructions for use,
• The warning label’s legibility, and
• The battery cycle count.
6.4 PRODUCT LIFE AND STORAGE