STANDARD COMPLIANCE
MESI mTABLET ABI is intended for use in the electromagnetic
environment specified below. The customer or the user of the
above listed models should assure that they are used in such an
environment.
Immunity Test IEC 60601
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Test Condition
±,8 kV Contact
± 2, 4, 8, 15 kV Air
±8 kV Contact
± 15 kV Air
± 1 kV
Signal input/output parts PORT
Electrical fast
transient/bursts
IEC 610004-4 (a)
± 1 kV
100 kHz Repetition frequency
Comment:
a) Not applicable to CLASS II ME EQUIPMENT and ME SYSTEMS.
b) The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to
6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz
to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to
21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz.
Mains power quality
should be that of a
professional healthcare
facility environment
and Home healthcare
environment.
Surges
IEC 61000-4-5 (b)
±2 kV line(s) to ground Not applicable.
Mains power quality
should be that of a
professional healthcare
facility environment
and Home healthcare
environment.
Conducted RF
induced by RF
fields
IEC 61000-4-6 (a)
3 Vrms
150 kHz - 80 MHz
6 Vrms in ISM bands between
150 kHz - 80 MHz c)
80% AM 1kHz
Mains power quality
should be that of a
professional healthcare
facility environment
and Home healthcare
environment.
3 Vrms
150 kHz - 80 MHz
6 Vrms in ISM
bands between 150
kHz - 80 MHz c)
80% AM 1 kHz