Line Tracer
21
4. Tehnical characteristics
4.1 Transmitter T10K
Batteries: 4 x AA size (1.5 V)
Low battery indicator: Built in
Operating temperature: 0
°C – +40 °C
Storage temperature: -30
°C – +60 °C
Weight: 280 g
Housing: ABS plastics
Dimensions: 80 mm x 50 mm x 150 mm
Overvoltages CAT III 300 V
Pollution degree 2
Automatic switching between generator and load mode according to input voltage.
Transmitter T10K in Generator Mode
DC, AC voltage: < 30 V or no voltage present on line
Operating frequency: 10.6 KHz modulated with 4 Hz
Maximum open circuit voltage: 6 V r.m.s.
Maximum short circuit current: 20 mA r.m.s.
Transmitter T10K in Load Mode
System voltage necessary for transmitting: 30 V ÷ 264 V, a.c.
System frequency: 45 Hz
÷ 65 Hz
Operating frequency: 10.6 KHz modulated with 4 Hz
Maximum average injected current: 1 A r.m.s.
4.2. Receiver R10K
Battery: 1 x 6LR61 size (9 V)
Low battery indicator: Built in
Operating temperature: 0
°C – +40 °C
Storage temperature: -30
°C – +60 °C
Weight: 140 g
Housing: ABS plastics
Dimensions: 45 mm x 45 mm x 210 mm
Built-in mechanical switch to select between capacitive and inductive mode.
Sensor: Built-in capacitive sensor for electric field tracing
and inductive sensor for magnetic field tracing
Selectivity: Input band-pass filter 10.6 KHz
Indicators: Audio - piezoelectric speaker (70 dB)
Visual - 10 level LED bargraph-style indicator
Sensitivity: LOW, MIDDLE, HIGH sensitivity keys, potentiometer for fine
adjustment of signal gain