EasyManua.ls Logo

METREL TeraOhm - Page 19

METREL TeraOhm
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TeraOhm 5kV
19
Measurement procedure:
- Connect the test leads to the instrument and to the tested object.
- Select INSULATION RESISTANCE function using the rotary switch knob.
- Press the START key and release it, continuous measurement starts to run.
- Wait until the test result has stabilized then press the START key again to stop the
measurement or until set timer runs out (if enabled).
- Wait for the object under test to be discharged.
- The result (see the right figure above) can optionally be saved pressing the MEM
key twice, see the chapter 6.1. Store, Recall and Clear Operation.
Legend of displayed symbols:
INSULATION RESISTANCE
Name of selected function
Off (Fil1, Fil2, Fil3)
Filter type enabled, see the chapter
5.3. Configuration
1000V
Set test voltage
U=1056V
Actual test voltage – measured value
I=0.04nA
Actual test current – measured value
>1.00T
Insulation Resistance – result
C=1.3nF
Capacitance of measured object
tm:00min 15s
Timer information – test duration
Bar
Analog presentation of result
Rmax=
Maximum value of result (only if timer
is enabled)
Rmin=
Minimum value of result (only if timer is
enabled)
Notes:
- If the timer is disabled then OFF is displayed instead of the timer value.
- During the measurement the timer information displays the time needed for the
completion of the measurement (tr) while after the completion the test duration (tm) is
displayed.
- A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a possible dangerous test voltage.
- Value of capacitance is measured during the final discharge of the measured object.
Set-up parameters for Insulation Resistance:
- Press the SELECT key, Set-up menu appears on
display, see the figure 10.
- Select parameter (line) to be set using the
and
keys;
- Adjust set parameter using the
and
keys. Skip
to the next sub-parameter by pressing the SELECT
key (if there are two or more sub-parameters) and
repeat the adjustment.
- Complete the set-up adjustments pressing either the
ESC key or START key (to run the measurement
directly) or changing the rotary switch position. Last
displayed settings are stored.
Fig. 10. Set-up menu in
Insulation Resistance
measurement

Related product manuals