EasyManuals Logo

METREL TeraOhm User Manual

METREL TeraOhm
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
TeraOhm 5kV
9
In case of a capacitive test object (long tested cable etc.), automatic discharge
of the object may not be done immediately after finishing the measurement –
“Please wait, discharging” message will be displayed.
HANDLING WITH CAPACITIVE LOADS
Note that 40 nF charged to 1 kV or 1.6 nF charged to 5 kV are hazardous live!
Never touch the measured object during the testing until it is totally
discharged.
Maximum external voltage between any two leads is 600 V (CAT III
environment).