Doc# 0296-0018 • REV R (July 2017) Page 33 of 61
Step 11. Use Input Test Cable (Prt. # 0215-0097-01). Connect the Calibraon Capacitor (Model 2947B-2)
between the synthesizer output channel 1 and the CHARGE CHANNEL input (J4) BNC on the Input Test
Cable.
Step 12. Repeat steps 5 through 8. Then disconnect the synthesizer and calibraon capacitor from the
TTS.
Step 13. Record the new Correcon Factor number on line 12 of the Check-O sheet.
Displacement DCCalibraon
Step 14. Use the Coaxial Cable (Standard BNC to BNC) and connect the synthesizer Channel 1 output to
the Displacement (J5) Input on the Input Test Cable.
Step 15. Set the synthesizer for 1V 0-peak sine wave at 100 Hz.
Step 16. Press the RUN buon to see the current input signal. The desired reading is 1000.00 ±.1%.
Step 17. If the reading is not between 999.00 and 1001.0, press the SPACE buon. The unit should
adjust itself within the limits.
Step 18. Record the Correcon Factor number on line 17 of the CHECK-OFF Sheet.
Step 19. Press the down arrow once. The DISPLACEMENT DC screen appears.
Step 20. Set the synthesizer for wave form DC and set the voltage to - 10 volts.
Step 21. Press the RUN buon to see what input signal the TTS currently reads. The desired reading is
-10000 +- .1%.
Step 22. If the reading is not between -9990 and -10010, press the SPACE buon. The unit should adjust
itself to within the limits.
Step 23. Record the Correcon Factor number on 21 of the CHECK- OFF Sheet.
Step 24. Disconnect the synthesizer from the displacement channel (J5).
Step 25. Press the down arrow once. The CALIBRATE C screen appears.
REF Channel No Integ Calibraon
Step 26. Mount the Transfer Standard Accelerometer (Model 2010M12) onto the TTS Reference Ac-
celerometer. Apply a lile vacuum grease on the surface of the Standard Accelerometer and the surface
of the Reference Accelerometer. (See Installaon instrucons for Installing Accelerometers in Chapter 3).
Step 27. Connect the Test Cable (Prt. #0215-0097-01) to the Transfer Standard Accelerometer and to the
Channel A (J1) input of the Input Test Cable.
Step 28. Get the sensivity value for the Transfer Standard Accelerometer from the Calibraon Docu-
ments that come with the standard accelerometer.
Step 29. Press the Cursor buon to move to the second line on the screen which reads REF NO INTEG
(Reference with no Integraon).
Step 30. Use the Le Arrow on the number keypad to remove the numbers. Use the number keypad to
enter the sensivity value from the Standard Acelerometer Documentaon.