Doc# 0296-0018 • REV R (July 2017) Page 35 of 61
Step 45. Press the Cursor buon to move to the second line on the screen which reads FREQ. CAL (Fre-
quency Calibraon). Enter the sensivity value that corresponds to the frequency to be set (3000 Hz).
You can get the sensivity value from the Standard Accelerometer documentaon.
• Press the RUN buon.
• Set the frequency to 3000 Hz and the amplitude (GIR) to 1.0 G.
• Press the SPACE buon to calibrate the frequency.
• The Correcon Factor should not be less than .85 or greater than 1.15.
• Record the Correcon Factor on line 29 of the CHECK-OFF Sheet.
• Press the down arrow buon. The CALIBRATE 4000 I screen appears.
Step 46. Press the Cursor buon to move to the second line on the screen which reads FREQ. CAL (Fre-
quency Calibraon). Enter the sensivity value that corresponds to the frequency to be set (4000 Hz).
You can get the sensivity value from the Standard Accelerometer documentaon.
• Press the RUN buon.
• Set the frequency to 4000 Hz and the amplitude (GIR) to 1.0 G.
• Press the SPACE buon to calibrate the frequency.
• The Correcon Factor should not be less than .85 or greater than 1.15.
• Record the Correcon Factor on line 30 of the CHECK-OFF Sheet.
• Press the down arrow buon. The CALIBRATE 5000 J screen appears.
Step 47. Press the Cursor buon to move to the second line on the screen which reads FREQ. CAL (Fre-
quency Calibraon). Enter the sensivity value that corresponds to the frequency to be set (5000 Hz).
You can get the sensivity value from the Standard Accelerometer documentaon.
• Press the RUN buon.
• Set the frequency to 5000 Hz and the amplitude (GIR) to 1.0 G.
• Press the SPACE buon to calibrate the frequency.
• The Correcon Factor should not be less than .85 or greater than 1.15.
• Record the Correcon Factor on line 33 of the CHECK-OFF Sheet.
• Press the down arrow buon. The CALIBRATE 6000 K screen appears.
Step 48. Press the Cursor buon to move to the second line on the screen which reads FREQ. CAL (Fre-
quency Calibraon). Enter the sensivity value that corresponds to the frequency to be set (6000 Hz).
You can get the sensivity value from the Standard Accelerometer documentaon.
• Press the RUN buon.
• Set the frequency to 6000 Hz and the amplitude (GIR) to 1.0 G.
• Press the SPACE buon to calibrate the frequency.
• The Correcon Factor should not be less than .85 or greater than 1.15.
• Record the Correcon Factor on line 35 of the CHECK-OFF Sheet.
• Press the down arrow buon. The CALIBRATE 7000 L screen appears.
Step 49. Press the Cursor buon to move to the second line on the screen which reads FREQ. CAL (Fre-
quency Calibraon). Enter the sensivity value that corresponds to the frequency to be set (7000 Hz).
You can get the sensivity value from the Standard Accelerometer documentaon.
• Press the RUN buon.
• Set the frequency to 7000 Hz and the amplitude (GIR) to 1.0 G.
• Press the SPACE buon to calibrate the frequency.
• The Correcon Factor should not be less than .85 or greater than 1.15.
• Record the Correcon Factor on line 37 of the CHECK-OFF Sheet.
• Press the down arrow buon. The CALIBRATE 8000 M screen appears.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrix HI 803 and is the answer not in the manual?

Metrix HI 803 Specifications

General IconGeneral
BrandMetrix
ModelHI 803
CategoryTest Equipment
LanguageEnglish

Related product manuals