12
Once the test has finished, the
display will show current test
(shown in black if more than one
maneuver has been performed,
the best maneuver will also be
displayed in blue) the MVV rate,
the % variation between
maneuvers, the breath rate and
the ATS quality warning for the
test session.
Select ‘Again’ to repeat the
maneuver, ‘Reject’ to reject the
current maneuver, ‘Results’ to
display a list of indices, the values
obtained, % predicted where
applicable and a quality statement
concerning the test session.
To meet the ATS quality criteria
for a good blow, the maneuver
should last 15 seconds with a
breath rate greater than 65
breaths per minute. The ATS
reproducibility criterion is two
maneuvers with a good blow and
the MVV variability between
maneuvers should not exceed
20%.
Note: The MVV test is an exhausting test. It should not be repeated
without a rest period. Some elderly or ill people cannot repeat this test
even after the rest period.