IM-LIPOTOWER Rev G 13
PREVENTIVE MAINTENANCE
The MicroAire LipoTower System requires the following preventive maintenance:
• The BIOFILTER ASSEMBLY should be replaced after approximately every 25 procedures, or sooner if
performance is reduced by clogging. If the BIOFILTER ASSEMBLY shows any moisture inside the lter or in the
tubing beyond the lter, it should be changed immediately.
• Replace with MicroAire part REF ASP-FLTR and reset the hour meter in the TOOLS window of the LCD screen.
• All MicroAire equipment should be inspected and tested periodically in accordance with the user facility’s
bioengineering policy. Such service should be documented within the bioengineering department, and on
the equipment (evidenced by an appropriate sticker).
WARNING: Repairs or alterations to MicroAire products made by anyone other than MicroAire or an authorized
MicroAire repair agent, will void the product’s warranty.
ENVIRONMENTAL PARAMETERS
OPERATING CONDITIONS
WARNING: If there is condensation present on the MicroAire LipoTower System, DO NOT use it. Wait for the unit to
come to room temperature and for condensation to evaporate before use.
WARNING: If the MicroAire LipoTower System is operated at high elevation or low atmospheric pressure, aspiration
performance may be adversely aected.
NOISE LEVEL EXPOSURE LIMITS
The A-weighted emission sound pressure level of the MicroAire LipoTower System operating under normal conditions
does not exceed 70 dB(A).
SHIPPING AND STORAGE CONDITIONS
SHIPPING: The materials and components used in the construction of this device were selected to insure that the
device could be shipped by any standard commercial method without special handling conditions.
STORAGE: Device should be stored in a designated, limited-access area that is well ventilated and provides protection
from dust, moisture, insects, vermin and temperature/humidity extremes.
REPAIR AND WARRANTY
MICROAIRE REPAIR SERVICE
Responsive service comes with every MicroAire product. If a problem should arise with your equipment, contact our
Customer Service Department at:
Telephone 800-722-0822
Fax 800-438-6309
Email inquiry@microaire.com
DO NOT disassemble or attempt to service the equipment. It can only be serviced by MicroAire, or an authorized
MicroAire repair agent. Unauthorized service will void the warranty.
Relative humidity limits
10%
91%
86
106
Atmospheric pressure limits
-18:r
49:
Temperature limits