IM-LIPOTOWER Rev G4
WARNING: Explosion hazard. Do not use in the presence of ammable anesthetics or oxygen.
WARNING: The MicroAire LipoTower System is heavy, weighing approximately 190 pounds. Use the proper equipment
to lift the system if required.
WARNING: Flat shelving capacity – 6.8 kg (15lb) maximum safe capacity. Do not exceed.
WARNING: Tumescent pole capacity - 3kg (6.6lb) maximum safe capacity per hook. Do not exceed.
WARNING: Do not transport cart with anything on shelves or hooks. Remove all items from cart before
unlocking casters.
WARNING: Do not modify any accessory. Failure to comply may result in patient and or operating room sta injury
and equipment damage.
WARNING: Use only MicroAire-approved accessories. Use of other unapproved accessories may result in increased
emissions or decreased immunity of the system and will void your warranty.
WARNING: This device will not, in and of itself, produce signicant weight reduction.
WARNING: This device should be used with extreme caution in patients with chronic medical conditions, such as
diabetes, heart, lung or circulatory system disease or obesity.
WARNING: The volume of blood loss and endogenous body uid loss may adversely aect intra and/or postoperative
hemodynamic stability and patient safety. The capability of providing adequate, timely replacement is essential for
patient safety.
WARNING: Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle
marked “Hospital Only” or “Hospital Grade.”
WARNING: Disconnecting the supply cord will isolate the system from the supply mains on all poles simultaneously.
WARNING: Inspect tubing and suction canisters for wear or damage before use.
WARNING: Risk of re. Use only MicroAire cables to connect to this device.
WARNING: Electric shock. Do not remove cover; refer servicing to qualied personnel only.
WARNING: Use care to ensure that there is no electromagnetic interference between this device and other
devices in use.
WARNING: Casters must be locked to prevent movement during normal use when a canister containing uid is
installed on the canister rack. Do not roll or tilt the cart in this condition because uid could spill onto the cart causing
a hazardous situation.
WARNING: ESD (Electromagnetic environment – guidance): Floors should be wood, concrete or ceramic tile. If oors
are covered with synthetic material, the relative humidity should be at least 30%. If ESD-related issues are experienced
during use, it may be necessary to use antistatic materials (e.g., anti-static mat) or to increase the relative humidity of
the surrounding environment.