5.2 Replacing Pipette Plunger Tips
It is recommended that when using the VP-ITC instrument on a daily basis, that the plunger tips
of the pipette be replaced approximately once every month or two. The tools for replacing syringe
plunger tips for the pipette injector consist of :
Removing old syringe plunger tip from pipette injector
You must first separate the Syringe Holder from the pipette
injector.
Place the pipette in its stand on the side of the instrument.
Open the Setup/Maintenance menu.
From the pipette maintenance group click on the Remove
Old Tip button.
Wait for the injector plunger to move up to its replacement
position (or home position), you will hear a beep when the
movement is complete.
Separate the injector from the Syringe Holder by turning the
Locking Collar counter clock-wise.