EasyManua.ls Logo

Microfluidics MICROFLUIDIZER LV1 Series - default chapter; Interaction Chamber; Model LV1 Microfluidizer Processor Data Sheet; Model Specifications

Default Icon
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model LV1 Microfluidizer Data Sheet
F-LV-1-5 8/3/2011 LV1 User Manual F-1
Model LV1 Microfluidizer Processor Data Sheet
Serial #:
Customer:
Intensifier
Ratio:
21:1 (Hydraulic Pressure: Process Pressure)
Process Piston Diameter:
0.437 in. (11.1 mm)
Stroke Length:
2.50 in. (63.5 mm)
Volume/Stroke:
0.375 in.
3
(6 ml)
Process Piston Seal:
UHMWPE (Ultra High Molecular Weight)
Gasket/O-rings:
Viton (standard)
Interaction Chamber
Style:
Serial #:
Model Specifications
Model Number
LV1
Approximate Process Discharge Flow
6 ml/stroke at full 2.5 inch per stroke
Maximum Product Processing Pressure
2069 bar (30,000 psi)
Product Inlet Pressure
Gravity feed
Inlet Reservoir Size
10 ml (0.61 cu-in)
Maximum Operating Hydraulic Pressure
124 bar (1,800 psi)
Maximum Hydraulic Oil Temperature
140°F (60°C)
Process Fluid Cooling Water
Required to cool from 161°F (72°C) to
70°F (21°C) [process temperature rises
approximately 3°F (1.7°C) per 1,000 psi
(69 bar) applied pressure]
Ice Water or Dry Ice used with optional cooling coil and tray
assembly
Intensifier Stroke Length
63.5 mm (2.50 inches)
Approximate Stroke Displacement
6 ml (.375 in
3
)
Approximate Strokes per Minute
Approximately 2 per minute (operator dependent) (see Chapter 3)