Ensure that neither the mains adapter nor the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket 6 in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.
7. Error messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
8. Safety, care, accuracy test and disposal
m
Safety and protection
Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical specifications» section.
The cuffs are sensitive and must be handled with care.
Only pump up the cuff once fitted.
Do not use this device if you think it is damaged or notice
anything unusual.
Never open this device.
Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
The measurement results given by this device is not a diag-
nosis. It is not replacing the need for the consultation of a physi-
cian, especially if not matching the patient’s symptoms. Do not
rely on the measurement result only, always consider other
potentially occurring symptoms and the patient’s feedback.
Calling a doctor or an ambulance is advised if needed.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.
m
Contra-indications
Do not use this device if the patient’s condition meets the following
contra-indications, to avoid inaccurate measurements or injuries.
The device is not intended for measuring blood pressure in
pediatric patients of age younger than 12 years old (children,
infant, or neonates).
Presence of significant cardiac arrhythmia during measurement
may interfere with blood pressure measurement and affect the
Error Description Potential cause and remedy
«ERR 1»
AS
Signal too weak The pulse signals on the cuff are
too weak. Re-position the cuff and
repeat the measurement.*
«ERR 2»
AR-B
Error signal During the measurement, error
signals were detected by the cuff,
caused for instance by movement
or muscle tension. Repeat the
measurement, keeping your arm
still.
«ERR 3»
AR-C
Abnormal cuff
pressure
An adequate pressure cannot be
generated in the cuff. A leak may
have occurred. Check that the cuff
is correctly connected and is not too
loose. Replace the batteries if
necessary. Repeat the measure-
ment.
«ERR 5» Abnormal result The measuring signals are inaccu-
rate and no result can therefore be
displayed. Read through the check-
list for performing reliable measure-
ments and then repeat the
measurement.*
«HI» Pulse or cuff
pressure too
high
The pressure in the cuff is too high
(over 299mmHg) OR the pulse is
too high (over 200 beats per
minute). Relax for 5 minutes and
repeat the measurement.*
«LO» Pulse too low The pulse is too low (less than 40
beats per minute). Repeat the
measurement.*

Table of Contents

Other manuals for Microlife BP B2 Basic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife BP B2 Basic and is the answer not in the manual?

Microlife BP B2 Basic Specifications

General IconGeneral
TypeAutomatic
Cuff size22 - 42 cm
Control typeButtons
Product colorWhite
Number of users1 user(s)
Measuring methodOscillometric
Memory registers30
Pulse rate range40 - 200
Speaking functionNo
Placement supportedUpper arm
Units of measurementmmHg
Mean arterial pressureYes
Display typeLCD
Battery typeAA
Battery voltage1.5 V
Battery technologyAlkaline
Battery life (CIPA standard)920 shots
Number of batteries supported4
Quantity per pack1 pc(s)
Pressure accuracy1 mmHg
Pressure measurement range20 - 280 mmHg