8
* Please immediately consult your doctor, if this or any other problem
occurs repeatedly.
8. Safety, care, accuracy test and disposal
Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
«ERR 3» No pressure
in the cuff
An adequate pressure cannot be gener-
ated in the cuff. A leak may have
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
If the 2-year warranty period has expired,
it is recommended to replace with a new
Microlife cuff available at your pharmacy.
«ERR 5» Abnormal
result
The measuring signals are inaccurate
and no result can therefore be displayed.
Read through the checklist for taking a
reliable measurement and then repeat
the measurement.*
«ERR 6» MAM Mode There were too many errors during the
measurement in MAM mode, making it
impossible to obtain a final result. Read
through the checklist for taking a reliable
measurement and then repeat the
measurement.*
Error Description Potential cause and remedy
«HI» Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
299 mmHg) OR the pulse is too high
(over 200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«LO» Pulse too low The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
Safety and protection
Error Description Potential cause and remedy

Table of Contents

Other manuals for Microlife BP W100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife BP W100 and is the answer not in the manual?

Microlife BP W100 Specifications

General IconGeneral
BrandMicrolife
ModelBP W100
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals