13Microlife BP W100 – MAM Wrist EN
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Dimensions: 80 x 70 x 70 mm
Cuff size: 13.5 – 21.5 cm (5.25 – 8.5 inches)
Measuring
procedure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range: SYS: 60 ~ 255 mmHg
DIA: 40 ~ 200 mmHg
Pulse: 40 – 200 beats per minute
Cuff pressure display
range: 0 – 299 mmHg
Resolution: 1 mmHg
Static accuracy: within ± 3 mmHg
Pulse accuracy: ± 5 % of the readout value
Voltage source: 2 x 1.5 V alkaline batteries; size AAA
Battery lifetime: approx. 460 measurements
(using new batteries)
IP Class: IP 20
Reference to
standards:
IEC 80601-2-30; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements,
whichever comes first
Accessories: 2 years or 5000
measurements, whichever comes first