54
Specications
Item
Subcomponent
Transmitter Sensor PDA
Model number
G7-T01
G7-T01A
G7-T01B
G7-S01
G7-S01A
G7-S01B
G7-P02
Operating temperature 5-40
℃(
41-104
℉)
Operating humidity 10-93%
(
non-condensing
)
Storage and transporta-
tion temperature
-20
℃
-60
℃
4
℃
-30
℃
-20
℃
-60
℃
Storage and transporta-
tion humidity
5-95%
(
non-condensing
)
Storage and transporta-
tion pressure
700hpa~1060hpa
Ingress protection level IPX7 IPX0
Use life 4 Years
G7-S01: 14 days
G7-S01A: 10 days
G7-S01B: 7 days
Shelf life: 1 year
4 Years
Detection range 2.0mmol/L-25.0 mmol/L
Measurement accuracy
When the glucose concentration >4.2mmol/l
(75mg/dL), the accuracy deviation of the sensor
does not exceed ±20%; when the glucose con-
centration ≤4.2 (75mg/dL), the accuracy deviation
does not exceed ±1mmol/l (18mg/dL).
Wireless frequency and
bandwidth
Frequency: 2.402GHz
~
2.48 GHz
Bandwidth: 1Mbps
Wireless modulation GFSK
Radiated power -2dBm