4
Precautions
CGMS readings should only be used as a reference
for the supplemental monitoring of diabetes mellitus
and should not be used as a basis for clinical diagno-
sis.
The CGMS should be completely removed before
Magnetic Resonance Imaging (MRI).
The CGMS contains many small parts that can be
dangerous if swallowed.
During rapid changes in blood glucose (more than
0.1mmol/L per minute), glucose levels measured in
interstitial uid by the CGMS may not be the same
as blood glucose levels. When blood glucose lev-
els drop rapidly, the sensor may produce a higher
reading than the blood glucose level; Conversely,
when blood glucose levels rise rapidly, the sensor
may produce a lower reading than the blood glucose
level. When experiencing symptoms of hypoglycemic
or hyperglycemic, the sensor readings are checked
by using a blood glucose meter for a ngertip blood

Other manuals for Microtech AiDEX