41
Using the built-in Blood
Glucose Meter (BGM)
The built-in blood glucose meter uses a chemical re-
agent (glucose oxidase, GOD) reaction to determine
blood glucose levels. After the test strip is placed into
the test strip port and a blood sample is applied, the
blood automatically wicks into test window. A transient
electrical current is generated, and this current is mea-
sured to determine the blood glucose level reading.
Blood sampling
Before testing, rst become familiar with how to collect
blood and then choose a clean and dry place to con-
duct the test.
Important
Prior to testing, use with either alcohol or soap wa-
ter to disinfect the sampling site. Use warm water
to increase blood flow if necessary. Dry your hands
and the sampling site, ensuring that there is no
soap residue remaining.

Other manuals for Microtech AiDEX