Contents:
Installation ..............1
Exploded View ........ 2
Special Tools:
none
Language of origin: English
Caution
The tray plate must be in place when using the cassette rack.
Failure to comply may result in equipment damage or personal injury.
Step 1: Install cassette rack onto tray plate.
A) Remove tray plate from autoclave.
B) Insert four cassette rack pegs into holes in tray plate.
Tray Plate
Pegs
Equipment Alert
Do not allow tray plate to contact water level sensor.
Failure to comply may result in an Error condition.
Step 2: Install cassette rack / tray plate.
A) Insert rack / plate into chamber as shown.
B) Slide rack / plate all the way to the back of the chamber.