EasyManua.ls Logo

Midmark M11 User Manual

Default Icon
2 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
English-1
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
TP200 Rev. A
MA8479i
M11 Cassette Racks [9A215]
003-2515-00 Rev. A (5/29/12)
Contents:
Installation ..............1
Exploded View ........ 2
Special Tools:
none
Language of origin: English
Caution
The tray plate must be in place when using the cassette rack.
Failure to comply may result in equipment damage or personal injury.
Step 1: Install cassette rack onto tray plate.
A) Remove tray plate from autoclave.
B) Insert four cassette rack pegs into holes in tray plate.
Tray Plate
Pegs
Equipment Alert
Do not allow tray plate to contact water level sensor.
Failure to comply may result in an Error condition.
Step 2: Install cassette rack / tray plate.
A) Insert rack / plate into chamber as shown.
B) Slide rack / plate all the way to the back of the chamber.
Style A
midmark.com

Other manuals for Midmark M11

Question and Answer IconNeed help?

Do you have a question about the Midmark M11 and is the answer not in the manual?

Midmark M11 Specifications

General IconGeneral
BrandMidmark
ModelM11
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals