To start a cycle...
A) Press the desired Cycle Button.
(Refer to ‘Cycle Parameters’ for times / temperatures).
B) Press the Start button.
Operating your M9 / M11 sterilizer...
Cycle Buttons
(Pre-set)
Cycle Buttons
(Programmable)
Step Action Description Step Action Description
1
Press desired button:
This selects the button that will be
programmed.
4
Press:
Program vent speed
2
Press:
Program sterilization temperature
Adjust:
Pressing “+” changes setting: FAST
Pressing “-” changes setting: SLOW
Adjust:
The “+” and “-” buttons adjust the
temperature by 1° increments.
5
Press:
Program dry time
(0 to 60 minutes)
3
Press:
Program sterilization time
Adjust:
The “+” and “-” buttons adjust the
time by 1-minute increments.
Adjust:
The “+” and “-” buttons adjust the
time by 1-minute increments.
6
Press:
The display will show the new cycle
parameters.
NOTE:
Pressing the STOP button during this procedure will abort the changes, and revert to the original settings.
The programmed settings are retained under Program # button <1> or <2>. Even if power is interrupted, or the unit is unplugged the setting will be retained.
Cycle Parameters
Unwrapped
270°F (132°C)
27.1 psi (186 kPa)
Sterilize: 3 min.
Dry: 30 min.*
• Instruments loose on a tray.
• Open glass or metal canisters.
• Tubing not used in surgical procedures.
• Loose items manufacturers recommend for exposure at 270°F (132°C).
Note: The sterility of unwrapped items is compromised on exposure to a non-sterile environment.
Pouches
270°F (132°C)
27.1 psi (186 kPa)
Sterilize: 5 min.
Dry: 30 min.*
• Pouched or loosely wrapped instruments.
• Multiple layers of instruments separated by fabric.
• Wrapped trays of loose instruments.
• Wrapped Cassettes.
• Tubing not used in surgical procedures.
• Wrapped items manufacturers recommend for exposure at 270°F (132°C)
Packs
250°F (121°C)
15 psi (104 kPa)
Sterilize: 30 min.
Dry: 30 min.*
• Textiles and surgical packs wrapped for sterilization.
• Items, except liquids, manufacturers recommend for exposure at 250°F (121°C) for 30 minutes.
Handpieces
270°F (132°C)
27.1 psi (186 kPa)
Sterilize: 6 min.
Dry: 30 min.*
• Dental handpieces (wrapped or unwrapped)
Note: Verify acceptability of sterilization parameters with handpiece manufacturer.
Programmable
User Dened
230°F to 275°F
(110°C to 135°C)
6 to 31 psi (41 to 214 kPa)
Sterilize: 3 to 90 min.
Dry: 0 to 60 min.
• Items appropriate for user’s dened parameters.
Caution
All material processed in these cycles must be validated by the user to ensure sterility of the
processed load.
Programmable cycles 1 & 2 are provided for those applications requiring sterilization parameters different than
the preset cycles. It is important to properly coordinate sterilization temperature with cycle time to achieve ster-
ilization. Sterilization temperature can be adjusted from 230° to 275°F (110° to 135°C). Permitted temperature
range for proper sterilization is 250° to 275°F (121° to 135°C). Temperatures set below 250°F (121°C) should
not be used for sterilization, unless required by the device manufacturer. Temperatures below 250°F (121°C) are
provided for disinfection only.
© Midmark Corporation 2012
not sterile.
004-1006-00 Rev. B 1/15/16
To Program Cycle Buttons...
* Refer to the User Guide for complete Cycle Operation procedures